20
Technical manual
|
X-am
®
2800
Maintenance
For inspection and maintenance, see e.g.:
– EN/IEC 60079-29-2 – Gas detectors - Selection, installation, use and
maintenance of detectors for flammable gases and oxygen
– EN 45544-4 – Electrical apparatus used for the direct detection and direct
concentration measurement of toxic gases and vapours - Part 4: Guide for
selection, installation, use and maintenance
– National regulations
6.2
Calibration intervals
Observe the relevant specifications in the DrägerSensor
®
& Gas Detectors Manual
and in the instructions for use/data sheets of the Dräger sensors installed.
Recommended calibration intervals for DrägerSensors:
National regulations may stipulate shorter intervals and must be applied.
6.3
Test gases
Test gas properties (e.g. relative humidity, concentration) can be found in the
relevant sensor data sheet.
The relative humidity of the test gas is not relevant for O
2
sensors.
Different test gases are used depending on the type of calibration.
6.4
Performing the bump test
The bump test can be performed as follows:
– Manual bump test
– Bump test with the bump test station
– Bump test with X-dock
The Dräger CC-Vision PC software can be used to configure whether a quick or
advanced bump test should be performed with the menu or in the bump test station.
Dräger recommends using the extended bump test for cross calibrations (see the
Dräger X-dock instructions for use).
For the CatEx sensor, we recommend methane as the test gas, including for cross
calibration, if methane is expected in the target application.
Prerequisites
DrägerSensor
®
Calibration interval
CatEx SR, XXS O
2
, XXS H
2
S LC, XXS
CO LC, XXS SO
2
, XXS NO
2
Every 6 months
1)
1) The recommended calibration interval can be extended to 12 months if it is ensured that an
advanced bump test (tolerance: max. ±20 %) is carried out and passed before use in safety-
relevant measurements. The advanced bump test checks the existing sensitivity.
CatEx SR, measured gas: H
2
Every 4 months
Other DrägerSensors
See the special data sheets for the
respective sensors.