Dräger Julian PCB CIO 8601051 Скачать руководство пользователя страница 1

Service 

    

Procedure

Field 

Date: 28 June 2002

© 2002 Draeger Medical, Inc.

Part Number: SP00249

Rev: -

Repair Instructions:

Julian PCB CIO 8601051

RETURN TO CD-ROM TABLE OF CONTENTS

RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS

Содержание Julian PCB CIO 8601051

Страница 1: ...e Procedure Field Date 28 June 2002 2002 Draeger Medical Inc Part Number SP00249 Rev Repair Instructions Julian PCB CIO 8601051 RETURN TO CD ROM TABLE OF CONTENTS RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS ...

Страница 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS ...

Страница 3: ...ing This procedure should only be performed by professional technicians with electronic soldering experience If not performed properly damage to the PCB may occur Warning Electrostatic discharge can damage electronic components When working on electronic equipment use an anti static mat and a wrist strap ESD Kit 1 Remove power to Julian 2 Remove upper left back panel 3 Locate and remove PCB CIO an...

Страница 4: ...lacement interval if necessary 5 Using side cutters cut the resistor leg from PIN 8 of the IC as close as possible to the IC see Figure 3 Using the needle nose pliers bend the resistor leg away from the IC so it is straight see Figure 4 Pull up resistor Figure 2 Pull up Resistor Location Figure 3 Cutting the Resistor Serial number label Figure 4 Soldering Buss Wire to Resistor RETURN TO CD ROM TAB...

Страница 5: ...r photo of completed rework see Figure 6 8 Locate the CIO PCB serial number label Change revision of PCB from 3 0 to 4 0 with permanent pen See Figure 2 9 Install new battery in CIO PCB Install reworked CIO PCB in accordance with Repair Instructions Electronics CIO PCB section 27 5 Replacing Battery on CIO PCB of the Julian service documentation Reconfigure machine to customer parameters 10 Reinst...

Страница 6: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS ...

Страница 7: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS ...

Страница 8: ...Draeger Medical Inc 3122 Commerce Drive Telford PA 18969 Tel 215 721 5402 800 543 5047 Fax 215 721 5784 Web www draegermedical com Printed in the U S A RETURN TO CD ROM TABLE OF CONTENTS RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS ...