Supplement
Fabius family SW 3.n
7
System overview
Symbols
Use of the external fresh-gas outlet with an addi-
tional switch:
When an automatic ventilation mode (
Volume
Control
,
Pressure Control
,
Pressure Support
,
SIMV/PS
) or the
ManSpont
ventilation mode is
used, set the switch on the external fresh-gas out-
let to the
COSY
position.
For further information, see "Using the external
fresh-gas outlet with an additional switch
(optional)" in the "Operation" chapter in the instruc-
tions for use.
Using the external fresh-gas outlet as a common
gas outlet:
When an automatic ventilation mode (
Volume
Control
,
Pressure Control
,
Pressure Support
,
SIMV/PS
) or the
ManSpont
ventilation mode is
used, connect the short fresh-gas hose to the
external fresh-gas outlet.
For further information, see "Using the external
fresh-gas outlet as a common gas outlet (optional)"
in the "Operation" chapter in the instructions for
use.
COSY