27
Operation
Configuring the pulse output
●
Press the key »
Param
« repeatedly until »
« is
displayed light on dark background.
●
Select the desired signal using
or
:
triggered mandatory stroke
or
mandatory stroke
or
alarm
Default setting:
mandatory stroke
Configuring RS 232 interface
To select data transmission rate (baudrate) and parity check
mode.
●
Press the key »
Param
« repeatedly until »
Baudrate
« is
displayed light on dark background.
●
Select the baudrate with the keys
or
:
9600
or
2400
or
1200
●
Press the key »
Param
« to select »Parity«.
●
Select parity check mode with the keys
or
:
NONE
or
EVEN
or
ODD
●
When using the printer select »
NONE
«.
Default settings:
Setting the printer
The HP ThinkJet must be configured as follows (cf. Owner's
Manual HP ThinkJet).
Baudrate:
9600
Parity:
NONE
Stop bits:
1 (fixed)
Data bit:
8 (fixed)
Baudrate:
same as Babylog 8000
Parity:
NONE
Data bits:
8
Handshake mode:
XON/XOFF
Alternate Control Sequence Mode