Dräger Alcotest 6510

Maintenance

6.3.1

Procedure for carrying out an accuracy check

1. From the Menu select 

"Accuracy check"

 and press the

 

OK

 - button.

{

The display shows the calibration check gas to be used.

{

"Next"

 is automatically highlighted at the bottom of the 

screen.

2. Press the 

OK

 - button.

{

After a short period, 

"Supply gas"

 is displayed.

3. Fit a clean dry mouthpiece to the device.
4. Connect the outlet from the check gas cylinder to the mouthpiece 

and allow gas to flow through the mouthpiece by pressing the val-
ve on top of the gas cylinder.

{

A Beep is heard, followed by an audible click, when the device 
has sampled the gas and 

"Remove gas, Analysing"

 is dis-

played momentarily.

5. Stop the flow of gas by releasing the valve at the top of the gas cy-

linder.

{

The Check value and the Result of the check are displayed 
and 

"Next"

 is highlighted at the bottom of the screen.

{

The result should be between 19 to 24.9 

μ

g/100 mL.

6. Record the result on the record sheet provided for that purpose. If 

the result lies outside this range contact the MBRS.

7. Press the 

OK

 - button to return to menu.

{

The display may show the recovery time counting down to 

 

zero.

8. Scroll up to 

"Back to measurement"

.

9. Press the 

OK

 - button.

{

The device returns to the Ready mode. 

7

Maintenance

The  operator may change the batteries as required (see page 13).

7.1

Testing

It is advisable that an accuracy check is carried periodically, i.e. on a 
monthly basis (see page 12).

7.2

Calibration

The device should be returned to the MBRS for recalibration at 5-7 
month intervals.

7.3

Storage

z

The Alcotest

®

 6510 consumes a small amount of current even 

when switched off. Therefore remove the batteries from the device 
during long periods without use.

z

The Device should not be stored under extremes of temperature.

z

Only dispose of as special waste in accordance with local waste 
disposal regulations. 

z

For more information about suitable waste disposal companies 
please contact the local authorities for environmental protection.

NOTICE

Accuracy checks may be carried out in either the 

"Specified 

"

 option or 

"Driver" 

option.

i

i

CAUTION

Used alkaline batteries must not be thrown into open fire, be 
recharged or opened by exerting force. Danger of explosion!

NOTICE

In case of long storage periods
Remove batteries from device, if the device is not used for 
more than 6 months to avoid damage to the device caused by 
leaking battery acid.

!

i

i

Содержание Alcotest 6510

Страница 1: ...Dr ger Alcotest 6510 Dual Limit Breath Alcohol Screening Device Approved by the Medical Bureau of Road Safety MBRS Instructions for Use i...

Страница 2: ......

Страница 3: ...4 5 Switching on and off 8 5 Performing the measurement 8 5 1 Requirements of the person to be tested 8 5 2 Automatic measurement 9 5 3 Error in delivering the breath sample 10 5 4 Removing the mouth...

Страница 4: ...sible for damage caused by non compliance with the recommendations given above The warranty and liability provisions of the terms of sale and delivery of Dr ger are likewise not modified by the recomm...

Страница 5: ...election of menu functions 3 OK button Confirms entries or selected functions Button to switch device on and off 4 Menu button DOWN MENU Entering menu functions Selection of menu functions 5 Socket no...

Страница 6: ...ompartment and remove cover 2 Observe correct polarity when inserting the batteries 3 Close battery cover 4 2 Batteries to be used 2 each 1 5 V Type Alkaline Mignon LR6 AA or rechargeable battery 2 1...

Страница 7: ...ping until the mouthpiece is correctly fitted to the device 2 01123840 eps With the grooved markings facing up press the mouth piece into mouthpiece receptacle until a click sound confirms correct en...

Страница 8: ...ions z A new mouthpiece is to be used for every person tested z Do not apply too high alcohol concentrations on to sensor e g by rinsing the mouth with high grade alcohols shortly before the measureme...

Страница 9: ...utton to confirm z 02623840 eps READY Test No 00140 After approximately 6 seconds READY is displayed and a short audible signal is given z At the same time the current test number is indicated A green...

Страница 10: ...d Use the M and buttons to choose person and the OK button to confirm The device is ready for measurement after 4 seconds If the subject exhales incorrectly e g by abruptly stopping or sucking at the...

Страница 11: ...on is highlighted z Activate the selected function by pressing OK button z Selecting Back to measurement and pressing the OK button brings the device into the measuring mode If no button is pressed wi...

Страница 12: ...may show the recovery time counting down to zero 8 Scroll up to Back to measurement 9 Press the OK button The device returns to the Ready mode 7 Maintenance The operator may change the batteries as re...

Страница 13: ...teries z Gently press onto the lock of the battery compartment and remove cover z Remove used batteries and in sert new ones consider cor rect polarity z Close battery cover 7 5 1 Required batteries...

Страница 14: ...es can be disposed of using special collecting and recycling facilities However as this device has not been registered for household usage it must not be disposed of through these means The device can...

Страница 15: ...ds Re select menu Device cannot be switched on Batteries discharged Replace batteries Insuffient volume Blow interruption Subject not blowing hard or evenly enough Instruct the subject to blow in the...

Страница 16: ......

Страница 17: ......

Страница 18: ......

Страница 19: ......

Страница 20: ...754 500 Dr ger Safety AG Co KGaA Edition 04 October 2011 Edition 01 April 2005 Subject to alteration Dr ger Safety AG Co KGaA Revalstra e 1 23560 L beck Germany Tel 49 451 882 0 Fax 49 451 882 20 80 w...

Отзывы: