Dräger 90 37 781 Скачать руководство пользователя страница 1

Breathing system heater

Fabius

®

 GS SW 1.3n

(90 37 781)

Fabius

®

 GS SW 2.n

(90 37 933)

Fabius

®

 GS SW 3.n 

(90 38 685)

Fabius

®

 Tiro SW 2.n

(90 38 132)

Fabius

®

 Tiro SW 3.n

(90 38 686)

Fabius

®

 GS 

premium

 SW 3.n (90 38 923)

Fabius

®

 plus SW 3.n

(90 39 054)

Supplement to Operator’s Instruction Manuals

Содержание 90 37 781

Страница 1: ...W 1 3n 90 37 781 Fabius GS SW 2 n 90 37 933 Fabius GS SW 3 n 90 38 685 Fabius Tiro SW 2 n 90 38 132 Fabius Tiro SW 3 n 90 38 686 Fabius GS premium SW 3 n 90 38 923 Fabius plus SW 3 n 90 39 054 Supplement to Operator s Instruction Manuals ...

Страница 2: ...4 Breathing System Heater Optional 4 Symbol Definition 5 General Warnings and Cautions 6 Installation of BSHPS Fabius GS 7 Installation of BSHPS Fabius Tiro 9 Operation of BSHPS 14 Switching Off BSHPS 16 Preparing the Compact Breathing System for Disinfection and Cleaning 16 BSHPS Technical Data 18 Daily Checkout 19 ...

Страница 3: ...facturers or an expert Not for use in areas of explosion hazard This apparatus is neither approved nor certified for use in areas where combustible or explosive gas mixtures are likely to occur Liability for proper function or damage The liability for the proper function of the apparatus is irrevocably transferred to the owner or operator to the extent that the apparatus is serviced or repaired by...

Страница 4: ... in conjunction with the Fabius GS and Fabius Tiro anaesthetic machines in order to reduce the condensation of moisture in the breathing system during low flow operation fresh gas flow below 1 l min This Instruction Manual supplement contains additions to the information contained in the Fabius GS or Fabius Tiro Operator s Instruction Manual ...

Страница 5: ...thing system heater Caution Refer to accompanying documents before operating equipment Caution Risk of electric shock do not remove cover Refer servicing to a DrägerService representative Degree of protection against electric shock Class 1 Type B Registration Mark Mains Applied Mains Power Fuse Heater Power DC Voltage AC Voltage C US ...

Страница 6: ...content falls below a specified minimum level the following undesirable reactions can occur regardless of the type of CO2 absorbent and the anaesthetic agent used e g Halothane Enflurane Isoflurane Sevoflurane or Desflurane reduced CO2 absorption formation of CO absorption and or decomposition of the inhalation anaesthetic agent increased heat generation in the absorber leading to higher breathing...

Страница 7: ...bius GS by sliding the mounting plate of the power supply through the opening of the rail 1 and then sliding it upward Tighten the knob 2 against the rail when the power supply is in the desired position 2 Route the power cable 3 for the heater below and behind the power supply and then up to the rear of the COSY arm Push the cable into the cable management cavity 4 on the bottom of the COSY arm 3...

Страница 8: ...ainst the rail when the power supply is in the desired position 2 Route the power cable for the heater through the trough 2 in Figure 2 on the rear of the cabinet At the right hand end of the trough route the cable into the cable management cavity 1 in Figure 3 on the bottom of the COSY arm Figure 3 BSHPS Installation on Machine with Right Mount COSY Step 2 3 Plug the connector at the end of the c...

Страница 9: ...abius Tiro by sliding the mounting plate of the power supply through the opening of the rail 1 and then sliding it upward Tighten the knob 2 against the rail when the power supply is in the desired position 2 Route the power cable 3 for the heater below and behind the power supply and then up to the rear of the COSY arm Push the cable into the cable management cavity 4 on the bottom of the COSY ar...

Страница 10: ...re 5 against the rail when the power supply is in the desired position 2 Route the power cable for the heater through the trough 2 in Figure 5 on the rear of the cabinet At the right hand end of the trough route the cable into the cable management cavity 1 in Figure 6 on the bottom of the COSY arm Figure 6 BSHPS Installation on Machine with Right Mount COSY Step 2 Trolley 3 Plug the connector at t...

Страница 11: ...of the power supply by removing the two screws 2 in Figure 7 Figure 8 BSHPS Installation on Wall Mounted Machine Step 2 2 Install the Dräger clamp onto the rear side of the power supply box using the two screws and lockwashers included with the clamp assembly Figure 8 Figure 9 BSHPS Installation on Wall Mounted Machine Step 3 3 Clamp the power supply onto the rear area of the left hand Dräger rail...

Страница 12: ...en up to the rear of the COSY arm 1 in Figure 10 Extra cable can be coiled around the clamp Push the cable into the cable management cavity on the bottom of the COSY arm 1 in Figure 11 5 Plug the connector at the end of the cable into the mating connector on the rear underside of the COSY 2 in Figure 11 Orient the connector so that the red mark on the connector plug aligns with the red mark on the...

Страница 13: ...e for the heater through the trough on the rear of the cabinet 1 in Figure 12 At the right hand end of the trough route the cable into the cable management cavity on the bottom of the COSY arm 2 in Figure 12 5 Plug the connector at the end of the cable into the mating connector on the rear underside of the COSY 3 in Figure 12 Orient the connector so that the red mark on the connector plug aligns w...

Страница 14: ... in use Warning When the moisture content falls below a specified minimum level the following undesirable reactions can occur regardless of the type of CO2 absorbent and the anaesthetic agent used e g Halothane Enflurane Isoflurane Sevoflurane or Desflurane reduced CO2 absorption formation of CO absorption and or decomposition of the inhalation anaesthetic agent increased heat generation in the ab...

Страница 15: ...ioning wall outlet Figure 16 BSHPS Connector HEATER POWER ON LED Is Not Illuminated If the HEATER POWER ON LED 2 in Figure 15 is not illuminated when the HEATER POWER switch 1 in Figure 15 is in the ON position and the MAINS POWER LED 1 in Figure 17 is illuminated check the fuse 3 in Figure 15 If it is blown replace it with the properly rated fuse Figure 17 BSHPS Back MAINS POWER LED Is Not Illumi...

Страница 16: ...nd Cleaning 1 Leave the Dräger Vapour s on the machine 2 Remove all breathing hoses 3 Remove the breathing bag 4 Remove the ventilation hose 5 Remove the fresh gas hose from the breathing system 6 Remove the anaesthetic scavenging hose 7 Detach the APL bypass and the Peep Pmax lines from the breathing system and from the side of the machine 8 Remove the flow sensor cable 9 Remove the O2 sensor cab...

Страница 17: ...iro Operator s Instruction Manual Cleaning disinfection of the mains adapter should be performed as described in the Fabius GS or Fabius Tiro Operator s Instruction Manual under Machine exterior Servicing Maintenance Please refer to the following Instruction Manuals Fabius GS SW 1 3n 90 37 781 Fabius GS SW 2 n 90 37 933 Fabius GS SW 3 n 90 38 685 Fabius Tiro SW 2 n 90 38 132 Fabius Tiro SW 3 n 90 ...

Страница 18: ...magnetic Compatibility EMC Conforming to EN 60601 1 2 IEC 60601 1 2 and ANSI AAMI IEC 60601 1 2 Electrical Safety Conformance IEC 60601 1 CAN CSA C22 2 No 601 1 M90 UL 60601 1 Environmental Temperature Operating 10 C to 35 C Storage 10 C to 70 C Humidity Operating 20 to 80 non condensing Storage 10 to 90 non condensing Barometric Pressure Operating 700 to 1060 mbar Storage 500 to 1060 mbar Classif...

Страница 19: ...ck power cord connections Ensure that the MAINS POWER LED is illuminated Ensure that the HEATER POWER switch is in the ON position Ensure that the HEATER POWER ON LED is illuminated Ensure that the underside of the COSY is warm approximately 35 40 C after 30 minutes have elapsed since the HEATER POWER ON LED was illuminated ...

Страница 20: ...r Use apply only to Fabius GS Fabius Tiro Fabius GS premium and Fabius plus with Serial No If no Serial No has been filled in by Dräger these Instructions for Use are provided for general information only and are not intended for use with any specific machine or device Directive 93 42 EEC Concerning Medical Devices Edition 4 2015 01 As of 2015 08 Dräger Medical GmbH changes to Drägerwerk AG Co KGa...

Отзывы: