![Drager Pac 5000 Скачать руководство пользователя страница 3](http://html1.mh-extra.com/html/drager-0/pac-5000/pac-5000_instructions-for-use_2527030003.webp)
5.9
Calibration and configuration
— For calibration or individual configuration connect Pac
5000 to a PC using the connecting cradle or the E-Cal
System. Calibration and configuration can be done with
installed software Pac Vision or CC Vision. A calibration
“due date” can be set using the operation timer (in days).
5.10 Adjustable operation timer (in days)
— Pac 5000 is equipped with an adjustable operation timer.
The operation timer can be used to set an individually
operation period e. g. to adjust a “calibration due date”,
an “inspection due date”, an “out of order date” etc.
— To adjust the operation timer connect Pac 5000 to a PC
using the connecting cradle or the E-Cal System. The
adjustment can be done with installed software Pac
Vision or CC Vision.
6
Alarms
6.1
Concentration Pre/Main Alarms
— The alarm will activate whenever the alarm thresholds A1
or A2 are exceeded. The instrument is equipped with a
vibrating alarm. It vibrates in parallel to these alarms.
— During an A1, the LED will blink and the alarm will sound.
— During an A2, the LED and alarm tone will repeat in a
double repeating pattern.
— The display will alternate between the measurement value
and “A1” or “A2”.
— The alarms may, according to the selected configuration,
be acknowledged or turned off. (See section 2.)
“Acknowledgeable”: alarms and LED can be
acknowledged by pressing [OK] for 1 second.
— “Latching”: The alarm will only deactivate when the
concentration falls under the alarm threshold and then
[OK] is pressed for 1 second.
— If the alarm is not latching, the alarm will deactivate as
soon as the concentration falls under the alarm threshold.
6.2
Battery pre/main alarms
— When the battery pre-alarm is activated, the audible alarm
sounds and the LED blinks, and the “low battery” icon
»
« flashes.
— To acknowledge the pre-alarm, push [OK] for 1 second.
— After the first battery pre-alarm, the battery will last for
approx. 1 further week and the “low battery” icon stays lit.
— When the battery main alarm is activated, the audible
alarm sounds in a repeating pattern of 2 repeating tones
and the LED blinks in the same pattern.
— The battery main alarm is not acknowledgeable; the
instrument will automatically turn off after approx. 1
minute.
— In case of a very low battery, the internal voltage monitor
could activate the LED’s.
6.3
Changing the battery
— Do not change the battery in explosion-hazard areas!
— The instrument contains a replaceable lithium ion battery.
— The battery is part of the Ex approval.
— Only the following battery types shall be used:
— Duracell DL 123 Lithium Type CR123A, 3 V
— ………
— ………
— Turn the instrument off.
— Unscrew the 4 screws from the back case.
— Open the front case and remove the depleted battery.
— Insert the new battery according to specified polarity
(+/–).
— Place front case back and fasten it by tightening the 4
screws of the back case.
6.4
Handling of exhausted batteries
— Caution:
— Never throw them into a fire!
— Never attempt to charge them!
— Never attempt to open them, danger of explosion!
— Dispose of exhausted batteries only as special waste in
accordance with local regulations.
— Spent batteries may be returned to Dräger for disposal.
6.5
Usable life alarm
— Before the end of the instrument’s usable life, a warning
period begins. During this period the remaining life time
flashes just after turning the instrument on, e. g. “30”/“d”.
— To acknowledge this message [OK] must be pressed for
approx. 1 second. After that, the instrument can be used
further.
— After the usable life has expired the text “0“/“d“ will
alternate in the display and cannot be acknowledged.
Pac 5000 will not longer measure and may be returned to
Dräger for recycling.
6.6
Instrument alarm
— The alarm and LED will be activated three times,
periodically.
— The [X] icon will be lit, with a 3 digit error code shown in
the display.
— If an error appears in the display see section 6.7 and if
necessary please contact DrägerService.
6.7
Trouble shooting errors
7
Technical Specifications
8
Sensor Specifications
Code
Cause
Remedies
100
Flash / EEprom write
fail
Contact DrägerService
102
AD system defect
Contact DrägerService
104
Flash check sum
wrong
Contact DrägerService
107
Self test failed
Contact DrägerService
210
Fresh air calibration
failed
Repeat operation
220
Span calibration failed Repeat operation
240
Bump test failed
Repeat operation or
calibrate instrument
Environmental Conditions
During
operation
–30 to 50
o
C
700 to 1300 hPa
5 to 95% relative humidity
Conditions
for storage
0 to 40
o
C
30 to 80% relative humidity
Ingress
protection
IP 65
Operating
times
2 years (typical at 25
o
C)
Battery life
(typical at
25
o
C)
24 hours of use per day,
2 minutes alarm per day:
CO, H
2
S: >8000 hours
O
2
: >3600 hours
Intensity of
alarm
typical 90 dBA at 30 cm
Dimensions
54 x 84 x 34 mm
Weight
120 g
Approvals
(only
approvals
printed on the
instrument
are valid)
CE Sign (89/336/EEC , 94/9/EC)
DMT 98 ATEX E021X
II 2 G EEx ia IIC T4, –40
o
C ..55
o
C.
I M1 EEx ia I, –40
o
C ..55
o
C
UL Class I, II, III Div I, Group A, B, C, D, F, G,
Temp Code T4, Exia
CSA Class I, Groups A, B, C, D, Exia Temp
Code T4
CO
H
2
S
O
2
Reproducibility
Zero point:
≤
±5 ppm
≤
±2 ppm
≤
±0.2 vol.-%
Sensitivity:
≤
±3 % of
measured value
≤
±5 % of
measured value
≤
±1 % of
measured value
Temperature Influence
Zero point:
≤
±5 ppm
≤
±2 ppm
≤
±0.2 vol.-%
Sensitivity:
≤
±0.4 % of
measured
value / K
≤
±5 % of
measured value
≤
±1 % of
measured value
Drift (20
o
C)
Zero point:
≤
±5 ppm/a
≤
±1 ppm/a
≤
±0.5 vol.-%/a
Sensitivity:
≤
±1 % of
measured
value / month
≤
±1 % of
measured
value / month
≤
±1 % of
measured
value / month
Please be aware of possible sensor cross sensitivities (see
Sensor Data Sheet XX XX XXX).
English
Содержание Pac 5000
Страница 1: ...D Pac 5000 CO H2S O2 Instructions for Use...