QUICK GUIDE BABYLOG® VN500

12 |

Getting started

 

Switching on Babylog VN500 and Infinity C500

Prerequisites:
–    Babylog VN500, Infinity C500,  

PS500 and GS500 are  

reprocessed and assembled  

ready for operation.

–  The mains power supply and  

  the gas supply are connected.
–   The Babylog VN500 toggle switch  

is set to On

1.  Press the key on Infinity C500.  

   The system is started. The start  

dialog is displayed.

Babylog VN500 provides you with  

two options:
–   Use settings of previous patient
 or
–  Admit new patient
–   If the Start dialog is closed using  

the X button, Babylog VN500 adopts  

the settings of the previous patient.

Ventilation unit,

left side:

Toggle switch

1.

Содержание Babylog VN500

Страница 1: ...Pocket Guide Babylog VN500 Software version 2 n D 24607 2009...

Страница 2: ...the Instructions for Use It is for informational purposes only Any use of the device requires full understanding and strict observation of the Instruction for Use The availability of functionality dep...

Страница 3: ...d also airway monitoring The Babylog VN500 ventilation unit is used with Dr ger Infinity C Series Medical Cockpits The Babylog VN500 ventilation unit is intended for use in different medical care area...

Страница 4: ...K GUIDE BABYLOG VN500 04 System overview Intensive care ventilator Control and display unit Infinity C500 Ventilation unit Gas supply unit GS500 Power supply unit PS500 Trolley with 4 brakes Babylog V...

Страница 5: ...view Cockpit Front Panel Alarm bar Touch screen AC power LED lights up when the device is connec ted to AC power Battery LED indicates battery status Audio paused button ON OFF button Rotary knob touc...

Страница 6: ...QUICK GUIDE BABYLOG VN500 06 System Overview Cockpit Back Panel VESA mounting holes Alarm bar USB ports Isolated network connectors Single link Digital Visual Interface DVI Serial communication ports...

Страница 7: ...07 System Overview Ventilation unit front panel Flap Expiration valve Water trap Inspiration valve Emergency air intake Operation display of ventilation Expiratory flow sensor Nebulizer port...

Страница 8: ...en Header bar Monitoring area with curves loops trends and measu red values Therapy bar with the therapy controls for the ventilation parameters of the active ventilation mode Main menu bar with butto...

Страница 9: ...Alarm status Alarm silence and display of remaining time Patient category System data e g state of charge of the batteries Alarms messages and instructions for the user Therapy status Therapy type ven...

Страница 10: ...to identification and the action required Colour Alarm priority Action required Red High Immediate action is necessary in order to avert an acute danger Yellow Medium Prompt action is necessary in ord...

Страница 11: ...ns Touch the control element The control element turns yellow Turn the rotary knob to the right or to the left to make the setting The value is displayed in the control element Confirm by pressing the...

Страница 12: ...pply and the gas supply are connected The Babylog VN500 toggle switch is set to On 1 Press the key on Infinity C500 The system is started The start dialog is displayed Babylog VN500 provides you with...

Страница 13: ...or No buttons The Next test button can be used to skip the test steps 5 5 1 1 The breathing circuit and the breathing gas humidifier can only be selected in standby mode 1 Touch the Start Standby but...

Страница 14: ...The check is only possible in standby mode Prerequisite The System check page is opened 1 Touch the Breathing circ check tab 2 Confirm with the rotary knob 3 When requested by Babylog VN500 in the in...

Страница 15: ...ettings by body height body weight Prerequisite The Weight function was configured and a new patient was admitted 1 Touch the button for the body weight height in ped patients 2 Set the weight body he...

Страница 16: ...e main menu bar The Start Standby page is displayed 2 Touch the Ventilation or the O2 Therapy button 3 Confirm with the rotary knob Babylog VN500 can switch between non invasive ventilation and tube v...

Страница 17: ...n in the main menu bar or 2 Touch the button in the therapy bar or 3 Touch the displayed ventilation mode in the header bar The Ventilation settings dialog window contains tabs for selecting the venti...

Страница 18: ...o confirm 4 Information is displayed in the message field e g when the setting limit of a parameter has been reached The ventilation modes can be combined with additional settings to optimize ventilat...

Страница 19: ...Connect a mask hood or nasal cannula for the O2 therapy 3 Touch the Start ventilation button and confirm with the rotary knob O2 therapy is switched on Babylog VN500 displays the main screen with the...

Страница 20: ...splay of alarms Acoustic alarm signals Alarm bar Alarm message field Blinking parameter field Babylog VN500 generates different alarm tone sequences to indicate alarms acoustically The acoustic alarm...

Страница 21: ...e Current alarms tab Acknowledging all alarm messages that are no longer active 4 Touch the Reset all button 5 Confirm with the rotary knob Setting alarm limits 1 Touch the Alarms button in the main m...

Страница 22: ...natal flow sensor is necessary During the device check and before use At least once a day After replacing the neonatal flow sensor After medication nebulization Recalibration is not necessary if the n...

Страница 23: ...ns for performing calibration The Confirm button is preselected Removing the neonatal flow sensor Remove the tube connector Put on a sterile glove Seal the neonatal flow sensor Performing calibration...

Страница 24: ...r gerwerk AG Co KGaA Branch Office P O Box 505108 Dubai United Arab Emirates Tel 971 4 4294 600 Fax 971 4 4294 699 contactuae draeger com REGION ASIA PACIFIC Draeger Singapore Pte Ltd 61 Science Park...

Отзывы: