APPENDIX A. SOFTWARE CONTROL
ASCII
E
r
\
0
\
0
value (4 byte )
Hex
0x45
0x72
0x00
0x00
data
The first two bytes are followed by two null characters whereafter the value is printed.
The value again needs to be typecasted to a little endian float.
In the case of a
read/write command that was successfully processed (”Ok”) the returned value will
contain the most current value of the feature which should be equal to the value that
is set. This does not hold when you exceed the setpoint limit, e.g. when you try to
set the RF power using the ”Pow” command to 999.0 the return value will be 100.0
since the power cannot be set to 999.0. All the available commands are listed below.
In the case of an error the four byte return value represents the error code, below you
can find what the meaning of the error codes are.
Commands (f
0
-f
2
) - Read/Write -
Case Sensitive !
Rem
remote mode (0 or 1)
Pha
phase between 0 and 100
Pow
power between 0 and 100
Swi
power switch (0 or 1)
Commands (f
0
-f
2
) - Read Only -
Case Sensitive !
Ref
Reflected power
Tra
Transmitted power
Tem
Cavity Temperature (measured by Pt100) in deg Celsius
Pfe
phase feedback enable (0 or 1)
Afe
Amplitude feedback enable (0 or 1)
Statera 15/100 version 1.1
16