www.drtrustusa.com
www.drtrustusa.com
10
11
10. REPLACEMENT OF BATTERIES
constant use.
·
The battery life is approx. 3000 readings one year and 90 minutes when in
·
Open the battery cover and remove the old battery.
·
Put in new battery and take care of the direction of the electrodes.
Cau on
1. When not used for 1 month, please take out the battery to prevent
2. According to local regulations, properly handle the waste batteries to
avoid pollution.
leakage. It is forbidden to put waste battery in the re.
11. CLEANING AND MAINTENANCE
Please keep the inner cavity of the sensors and probes clean, otherwise it
may affect the measurement accuracy.
Cleaning Methods
with a little medical alcohol or water to wipe the dirt.
·
The device only needs routine cleaning, and period or times can be
decided according to user’s need.
·
For surface cleaning, use a clean soft cloth or cotton swab stick moistened
·
For probe and sensor inner cavity cleaning, use clean soft cloth dipped
with a little medical alcohol to gently wipe the probe at the top of the
inner cavity or sensors, and do not use until the alcohol completely
our customer support for further queries.
evaporates.
·
No Special maintenance in needed to the use process. Please contact to
Monitor Disinfection
·
Switch off the monitor and disconnect it from the power line.
disinfectant solution.
·
Wipe the display screen using a soft, clean cloth dampened with the
·
Wipe the exterior surface of the equipment using a soft cloth dampened
with the disinfectant solution.
·
Wipe off the disinfectant solution with a dry cloth after disinfection if
necessary
·
Dry the monitor for at least 30 minutes in a ventilated and cool place.
Warning
·
Please keep the inner cavity of the sensors and probes clean, otherwise it
will affect the measuring accuracy.
·
If you spill liquid on the probe sensor or LCD, or they are accidentally
immersed in liquid, contact your service personnel or service engineer.
·
Use a soft, dry cloth to clean the thermometer body and LCD display.
and disconnected from the power line.
·
Before cleaning or disinfection make sure that the device is switched off
or any other liquid.
·
After cleaning with isopropyl alcohol, allow the thermometer to air dry.
·
Do not use abrasive cleaners. Never submerge this thermometer in water
12. TROUBLESHOOTING
Table 4