4.4.2 Changing the fuses
♦
Ceiling/ wall/ stand model
Power supply with 24V AC/DC
In the case of designs for power supply with 24V
AC/DC the necessary fuses are provided by the
customer.
♦
Ceiling/ wall model
Power supply with 230V/120V AC -
built in toroid transformer
The fuses in the lamp housing (in the toroid trans-
former design) prevent the transformer burning
through in the event of a short circuit.
Visible type 5x20/2,00A/250V/T for 230 V
Visible type 5x20/4,00A/250V/T for 120 V
If the bulb is no longer on, check the bulb first, then
the fuses.
To change the fuses proceed as follows:
•
Turn off the light.
•
Turn the black fuse holder
1
1/4 of a rotation
anticlockwise with a screwdriver.
•
Change the fuses.
•
Insert the fuse holder under slight pressure and
rotate in a clockwise direction (bayonet catch).
Pay attention to the instructions from the local
installer.
1
2
1
♦
Stand model
Power supply with 230V AC
The fuses are in the counter balance
2
as shown in
the figure.
Visible type 5x20/2,00A/250V/T for 230 V
To change the fuses proceed as follows:
•
Turn off the light.
•
Turn the black fuse holder
1
1/4 of a rotation
anticlockwise with a screwdriver.
•
Change the fuses.
•
Insert the fuse holder under slight pressure and
rotate in a clockwise direction (bayonet catch).