Mach M5
Dr. Mach
Lamps and Engineering
5.2 Electrical data
Mach M5 DF / M5 DF/H
Power consumption
231 W
Voltage AC/DC
22,8 V
Current
10,1 A
Halogen bulb
22,8 V / 77 W
Frequency
50/60 Hz
Degree of protection
Type B
Class of protection
I.
5.3 Environemental conditions
Operation
Min. Max.
Tempe10°C
+40°C
Relative athmospheric humidity
30 %
75 %
Air pressure
700 hPa
1060 hPa
Transport / storage
Min. Max.
Temperature -10°C
+50°C
Relative athmospheric humidity
20 %
90 %
Air pressure
700 hPa
1060 hPa
5.4 General remarks
When using more than one OT-lamp at the same time (OT-lamp combinations), please make sure that
the total radiation intensity in field is less than 1000 W/m
2
, to avoid further temperature increase in the
wound field.
When installing an OT-lamp, its fail-safety must be guaranteed according to DIN VDE 0100-710 (former
DIN VDE 0107).
Protective conductor
59270001
Edition 06
28.09.2004 / Bak page 20/26