NT: nn min

This is the time (minutes) remaining until next testing.
 

W: nnn

This is the step motor setting value to control how many steps motor will turn to pump water into the 
reaction chamber (unit: steps).
 

AutoMode: n

Automatic testing mode (1 = ON, 0 = OFF).
  

P: STANDBY

Current system status.
 

SYS: READY

Sublevel control status (For support staff use only).
 

L.Time: MM/DD hh:mm

Date and time of the last testing performed.
 

C.Time: MM/DD hh:mm

Date and time of the current testing performing.

Page 7

KHG System Setup

Auto mode:

Default value is 1. 1 means KHG is in automatic mode. Any other value will disable 
automatic mode.

Check intervals (Min):

This is to set the minutes intervals between each KHG test. Unit is minutes. We 
recommend setting the intervals to 180 minutes.  You can shorten the testing 
interval time, but that will also use up KH reagent solution faster. 

No_SW_ERR :

This is to set the alarm if the number of steps of the step motor has turned, but 
still not enough water has been pumped into the reaction chamber. Default value 
is from 140 to 160. The unit is number of steps for the step motor. This is useful 
to determine if there were blockage in the SW feeder tube. Depending on how 
far your KHG is placed from the tank water source and how much head there 
are to pump water into the reaction chamber. You can adjust this value to best 
suit your environment. However, caution must be taken if you deviate from the 
recommended range too far, it may result in excessive water being pumped out of 
your system.

1.

3.

2.

(Figure 11)

User Manual

Содержание KHG

Страница 1: ...WWW KHGUARDIAN CO M V8 0C Profashional Edition 2017 01 09...

Страница 2: ......

Страница 3: ......

Страница 4: ...ction Main Display KHG System Setup The Meaning of Log Field Error Message and Troubleshooting Additional User Instructions The KHG exterior features Main Control of KHG P1 P5 P1 P5 P7 P10 P10 P12 P1...

Страница 5: ...bicarbonate solution should be added to the aquarium with the reference of the desired dKH target value aquarium total water volume and current testing value Then the controller dose in the solution...

Страница 6: ...e sure the placement of KHG is level on the X Y and Z axis Complete levelling is crucial as to keep the sodium bicarbonate solution from dripping on the optical and electronical components before ente...

Страница 7: ...eagent solution bottle Please insert the pH probe into pH 7 0 calibration solution of at least 1 5 cm deep Press and hold number 2 and number 4 buttons simultaneously for 3 second until you hear a bee...

Страница 8: ...many times as out of the inlet tube Please click on the KS De Gas on your browser to purge air from your KHG reagent solution tube Please look into the observation hole Figure 6 until you see reagent...

Страница 9: ...t be performed every time you intend to move the KHG It is crucial to purge reaction chamber to prevent solutions damaging the optical lens and other electronic components Failure to purge chamber cou...

Страница 10: ...not use unless being instructed by a qualified KHG support staff ENG Mode OFF Advanced function only Do not use unless being instructed by a qualified KHG support staff Test KH doser Press to test if...

Страница 11: ...between each KHG test Unit is minutes We recommend setting the intervals to 180 minutes You can shorten the testing interval time but that will also use up KH reagent solution faster No_SW_ERR This is...

Страница 12: ...ntroller KHG not only monitors your tank alkalinity but with KH auto correction KHG will also actively maintain your desired dKH level Regardless if you normally maintain your tank alkalinity by kalkw...

Страница 13: ...Save all settings above to the memory KHG will force reboot after new settings have been saved New PW Please enter new password Password allows maximum 8 characters include spaces numbers letters or...

Страница 14: ...e Please gently open the Reaction Chamber and make sure the white water level sensor is freely movable up and down Sensor may be malfunctioned Please contact your dealer eg 10 07 15 07 49 W 131 0 AK 3...

Страница 15: ...ensor may be malfunctioned Please contact your dealer KH ERR KH value high error A B C D E F 4 This error message indicates the tested value is 1 5dKH larger than the KH auto correction setting System...

Страница 16: ...bration procedure If all settings are correct please contact your dealer Page12 4 Additional User Instructions 1 2 3 Upon initial assembly and installation please make sure to perform 5 five dKH readi...

Страница 17: ...the factory supplied KHG reagents However the lack of preservatives may result in black color bacteria substances growing in the tubing From the test result of using KHG in the last 4 years these sub...

Страница 18: ...mixing 65 grams of sodium bicarbonate powder per one liter of water Please keep the error tolerance to within 5 We would strongly recommend to keep KHG in a protected container away from saltwater sp...

Страница 19: ...Press and hold buttons 2 4 for 3 seconds To KS_DS_Gas Press and hold buttons 1 2 for 3 seconds To SW_Tube De_Gas Press and hold buttons 1 3 for 3 seconds To Purge Chamber Press and hold buttons 3 4 fo...

Страница 20: ...Contact SALES KHGUARDIAN COM SUPPORT KHGUARDIAN COM...

Отзывы: