18. Trouble-Shooting
Dr. A. Kuntze
I
Gutes Wasser mit System
Robert-Bosch-Str. 7a
40668 Meerbusch
+49 (0) 2150-7066-0 / Fax -60
www.kuntze.com
92
Change the SD card, or remove some of its contents, or select the mode RING to
overwrite older data.
Popup message
Cause
Suggested remedies
Set point 2 not available
Set point 1 was set to 3-
point controller
No setting possible, since
set point 1 is set to 3-point
controller
If you have selected a 3-point controller for set point 1, then both control relays are
used for this set point, one to open the actor and one to close it, so you cannot use
set point 2.
18.9. Reset
As a last resort it is sometimes necessary to restore factory settings. With the RESET
function you erase all settings by operators. Activated add-ons will of course remain
activated.
With the Reset function, the system is set to a defined condition. User settings that
might severely impede an evaluation of the measured values, such as a calibration
gone wrong, are erased.