Lower limit
/ upper limit
You can define an upper and a lower limit. If the measured value is higher than the
upper limit or lower than the lower limit, an alarm is issued via the alarm relay
– if you
have specified the limits as alarm actions.
13.2. Alarm action
Select the symbol ALARM in the main menu.
Select the symbol ACTION.
A list of all possible actions appears.
By activating the check boxes you can define actions as alarm actions. If a box is
marked with a tick, the action or event will cause the alarm relay to switch. An empty
box means that the action or event does not cause the alarm relay to switch.
Overrange refers to both inputs, the message and alarm will occur if either input is
affected.
Note
If an action concerns an add-on that you have not activated, the
box is crossed out and cannot be selected.
The following events always cause the alarm relay to switch:
Communication error
Unknown measuring module