10.
Test menu
The menu TEST is only accessible in the MAN Mode. In all other modes, the symbol
is not shown. The test menu always shows a list of the maximum available analog
and digital outputs. Naturally, you can operate only those outputs that are activated in
your instrument. In the entry-level version this means only relay 3.
Set the mode to manual. Select the symbol TEST in
the submenu IN/OUT.
In the submenu TEST you can switch all available relays and define mA values for
the analog outputs. To switch a relay, select ON from its drop-down list.
To check an analog output, select its settings to open the keyboard. Set the output to
a value between 0.00 and 22.00 mA. This value is then sent out via the selected
analog output.
Note
When you leave the menu, all relays and all analog outputs are
reset to their previous settings.