6.
Analog input
– pH or ORP measurement
Select the symbol IN/OUT in the main menu and then
the symbol ANALOG. A submenu opens, showing
three drop-down lists, or two, if ORP has been
selected.
In the submenu ANALOG you can switch from pH to
ORP and back, activate a suitable averaging level, and
define the connection (for pH-differential sensors).
Measure
To select the parameter, select pH or mV (ORP).
Averaging
To activate the averaging function, select:
OFF : no averaging
low: mean of 5 values
medium: mean of 10 values
high: mean of 20 values
Connection
Select Triax if you want to connect a pH-differential sensor with a high-impedance
reference and a third electrode as ground potential. Select Koax for common pH
sensors with or without integrated temperature sensor. Mind that the sensor and
sensor connection must fit the setting.