DATA LOGGER SOUND LEVEL METER
26
POWER-UP OPTIONS
Med= 50
Auto=
Hi = 80
30
REC
Lo = 30
100 dB
130 dB
130 dB
80 dB
LEVEL
SETUP
MIN
CLOCK
MAX
INTV
SLOW
FAST
A
C
O
U
T
P
U
T
D
C
9
V
C
A
L
R
S
-2
3
2
VI. CALIBRATION PROCEDURES
§
Using a standard Acoustic Calibrator
( 94dB , 1KHz Sine wave )
(1)
Make the following switch settings.
Display : dBA
Time weighting : FAST
Measurement mode : MAX MIN Mode function disable.
Level range : 50 to 100dB
(2)
Insert the microphone housing carefully into the insertion hole of
the calibrator.
(3)
Turn on the switch of calibrator and adjust the CAL potentiometer
of the unit . The level display will indicate the desired level.
All products are well calibrated before shipment.
Recommended Recalibration cycle: 1 year.
Acoustic Calibrator
Screwdriver
Содержание 5020-0322