BFP SYSTEM :
Quick Starting Guide
4.6. Recording and Data Saving
(Acquisition Tab)
There is 3 operating mode for the system :
4.6.1. LIVE :
This mode is acquiring images, but does not save the file on the computer. It allows to setup your
experiment, and test the system without saving useless data on your computer.
4.6.2. RECORD :
This is the mode you should use when doing single experiment recording. First step is to define the the
file name and target path in the
Saving Options
.
By default, we record the average values of all ROI, Digital I/O if any have been set, and the key event
that create a timestamp everytime user presses the F1, F2, F3 or F4 key. In some computers, it is
required to press Alt+F1 to have this functionalty working. In future, software will allow more
configurable Event keys. Video files are not saved by default. If Save Video is selected, then a separate
video file is created for each feed camera/excitation visible in the interface for the configuration that
have been set.
* Please note that if no ROI have been set, nothing will be recorded.