![Doran Scales DS9100 Скачать руководство пользователя страница 11](http://html1.mh-extra.com/html/doran-scales/ds9100/ds9100_instruction-manual_2521654011.webp)
7
Reweigh
To reinitiate the weighing process, press REWEIGH. This will allow a weighment
to be reinitiated without the patient having to get off the scale platform.
BMI
Once a weight is held, the height of the patient can be entered using the full
keypad by pressing the BMI button.
Display Annunciators
Motion
When the scale platform senses motion, MOTION will be indicated. Once motion
ceases, a stable hold reading will be displayed.
Center of Zero (
0
)
Indicates that the scale has an accurate zero reading.
Stable Hold
Indicates a stable weight has been achieved and held on the display. This
indicates an accurate weight has been achieved.
Average Hold
Indicates that the weight held on the display is a result of averaging several
weights due to excessive motion on the scale platform. The Average Hold weight
may be slightly less accurate than a Stable Hold weight.
Displaying an Average Hold weight instead of a Stable Hold weight will speed the
weighment process of patients that cannot stand still. The Average Hold weight
will differ from the patient’s actual weight based upon the severity of the motion
caused by the patient while on the scale. Although this weight may be less
accurate than a Stable Hold weight, the deviation from the patient’s actual weight
will be minimal.
The time required for the Average Hold to engage is based upon the Average
Aperture parameter. See the Parameter Setup section for configuration
instructions.
Last Weight
Last Weight is indicates when a previous weight is being displayed.
Battery Indicator
The battery indicator displays the remaining battery life available. See battery
operation section for detailed information.
NET
Indicates a net weight is displayed when a tare weight is active.