HOLD function status
Fixed value
Fixed bar graph
①
②
3) Position the sensing plate at ionizer to measure ion balance. Please approach the sensing
plate on ionizer to measure ion balance as much as distance between ionizer and object.
Ion balance value could be different according to the distance.
4) When measuring on Electrostatic mode, pull down the plate rotary button(
①
) on both
side of head. Please fold the sensing plate(
②
) up for exposing the sensing hole.
13
7.6 HOLD function [Electrostatic mode/Ion balance mode]
When the value on DISPLAY is hard to be checked, this function holds the value on DISPLAY.
So it is convenient to check the value even after changing the position.
1) Follow the operating instruction for measurement.
2) Press [HOLD] button on main body once when to measure the value.
3) When positioning the product with measurable distance, the value is fixed and “HOLD” is
displayed on LCD.
7. Operating Instruction
DONG IL TECHNOLOGY LTD
When “HHH” or “LLL” is displayed, the level of static electricity is over measurable
range.Please stop measuring. Otherwise, integral sensor can be damaged.
※ HHH : over (+)voltage measurable range[+300V], LLL : over (-)voltage measurable
range[-300V]
Warning!!