Measuring range:
Pressure dewpoint in °C
td
UDM 300:
-80...+20°C
td
scaling 4...20 mA =^ -110...+20°C
td
(special scaling upon request)
Pressure range:
-1...+50 bar standard
350 bar high pressure (special version)
Power supply:
24 VD
C
230 VAC, 50...60 Hz for UDM 300 - KPL
complete instrument
Accuracy:
±0.5°C
td
(-10...+50°C
td
)
typical ±2°C
td
at -40°C
td
Output:
4...20 mA in 2-wire technology
Protection class:
IP65
CE conformity to:
DIN EN 61326
Operating temperature:
-20...+70°C
Storage temperature:
-40...+80°C
Load for analogue output: <500 Ohm
Screw in thread:
G 1/2" stainless steel
Material of housing:
polycarbonate
Sensor protection:
sinter filter 50 mm stainless steel
8
The UDM 300 complies with the EMC
standard according to the industrial stan-
dard 61326. We recommend the use of a
shielded connection cable in particular
with long signal transmissions and "harsh
industrial conditions". When using a mea-
suring chamber, earth one side of the
shield in the control cabinet. For direct
installation without measuring chamber,
connect the shielding as short as possible
to clamp 4 (earth) in connection plug on
UDM 300.
TECHNICAL DATA
Содержание UDM 300
Страница 14: ...UDM 300 Komplettger t mit DS 52 14...
Страница 30: ...UDM 300 COMPLETE INSTRUMENT with DS 52 14...
Страница 31: ...UDM 300 COMPLETE INSTRUMENT with DS 52 15...