2-2 21Vol% Adjustment 

 

 

  (1) Set to maintenance mode

 (MNT1 or MNT2).

 

 

 

 

 

 

 

 

 

 

 

 

 

 

(2) Release the password lock. (Refer to 

1-2 Releasing the Password Lock

.) 

 

(3) Press the SPAN switch. 

 

 

 

 

 

 

 

(4) Let the calibration gas through. 

 

(5) Use the UP/DOWN switch to adjust the concentration indication on the screen to that of the calibration 

gas concentration.   

 

(6) When the adjustment is completed, press the SPAN switch again. The SPAN indication in the bottom 

right corner of the LCD screen will then disappear, and the 21vol% adjustment is complete. 

 

(7) Remove the calibration gas and confirm that no alarm is displayed(

 ALARM 1 , ALARM 2

 is not shown). Then, 

restore to the normal mode by switching the maintenance switch to the center. 

Pull back to open 

 

Left: Maintenance 1 (MNT1) 

Center: Normal mode 

Right: Maintenance 2 (MNT2)

 

TROUBLE lamp(yellow

blinks.

MNT1 or MNT2 is displayed at the
bottom of the LCD screen. 

The “__” indication and the concentration 
value of the Oxygen blink alternately.

 

SPAN is displayed at bottom of the
LCD screen

SPAN switch 

Oxygen sensor unit: COS-7

 

Содержание PS-7

Страница 1: ... 5300 Fax www dodtec com Operate this unit only after reading and fully understanding the content of this manual This Operation Manual was written for administrator use Instructions for the basic operations of this unit are provided separately in the standard Operation Manual Refer to the Operation Manual when necessary For Administrators ...

Страница 2: ...ettings Refer to 2 4 Changing the Alarm Settings 1 2 Releasing the Password Lock 1 Press these key switches in the following order 1 ZERO switch 2 UP switch 3 Special command switch 4 DOWN switch 2 The password lock is released when the POWER lamp green blinks once 3 The password lock is automatically reactivated 10 minutes after it has been released 1 2 3 4 POWER lamp green Blinks once Warning Im...

Страница 3: ...creen will go off and the span adjustment will be completed 7 Remove the calibration gas check that the detected gas concentration is under that of the alarm value and restore to the normal mode by switching the maintenance switch to the center Toxic gas sensor unit CDS 7 Combustible gas sensor unit CHS 7 Pull back to open Left Maintenance 1 MNT1 Center Normal mode Right Maintenance 2 MNT2 TROUBLE...

Страница 4: ...ttom right corner of the LCD screen will then disappear and the 21vol adjustment is complete 7 Remove the calibration gas and confirm that no alarm is displayed ALARM 1 ALARM 2 is not shown Then restore to the normal mode by switching the maintenance switch to the center Pull back to open Left Maintenance 1 MNT1 Center Normal mode Right Maintenance 2 MNT2 TROUBLE lamp yellow blinks MNT1 or MNT2 is...

Страница 5: ...ock Refer to 1 2 Releasing the Password Lock 2 Press the UP DOWN switch U or V until the setting you wish to change LCD Screen Indication in the table below is displayed at the bottom right of the LCD display 3 Then press the special command switch and the setting to be changed LCD Screen Indication in the table below and will flash alternately Special command switch Setting to be changed UP DOWN ...

Страница 6: ...ue rounded to the percent zs 5 zs 2 0 to 30 H H L L H L Alarm mode 1st upper limit 2nd upper limit warning 1st lower limit 2nd lower limit warning 1st upper limit 2nd lower limit warning H H L L H H L L H L Con Pyroletic converter failure alarm 0 Off 1 On Con 0 Con 0 0 or 1 CG Calibration gas concentration For our maintenance purposes only CG 40 CG 84 nEt The existence of a DeviceNet unit 0 Non ex...

Страница 7: ...72 72 F S 0 to 100 F S Toxic CDS 7 Comb CHS 7 A2 20 20 F S A2 Oxygen COS 7 A2 76 76 F S 0 to 100 F S Pressing the AL switch changes to 1st level Æ 2nd levelÆ Normal display The A2 value should be greater than the A1 value in any case 3 Press the special command switch The setting to be changed and the indication will then flash alternately 4 Change the value by using the UP DOWN key U and V When y...

Страница 8: ...vel alarm ALARM 1 ALARM 2 is displayed Alarm contact ZA1 and ZA2 are activated Alarm value is set by entering A1 0ppm Full Scale 2 nd level gas alarm ALARM 1 ALARM 2 is displayed Alarm contacts ZA1 and ZA2 are activated Alarm value is set by entering A2 1 st level gas alarm ALARM 1 is displayed Alarm contact ZA1 is activated Alarm value is set by entering A1 Upper alarm ALARM 2 is displayed Alarm ...

Страница 9: ...able Contact the following address for ordering information Manufacturer New Cosmos Electric Co Ltd 2 5 4 Mitsuya naka Yodogawa ku Osaka 532 0036 Japan Phone 81 6 6309 1505 Fax 81 6 6309 1514 Distributor DOD Technologies Inc 740 McArdle Drive Unit C Crystal Lake IL 60014 Phone 815 788 5200 Fax 815 788 5200 www dodtec com ...

Отзывы: