Configuring Test Plans
4-64
72A-1898 Rev. A 11/01
Monday
, November 26, 2001 1:09 pm
Trigger Input Tab
Trigger inputs can be configured for all tests, except Continuity,
Self-Diag., Capacitance and Slow Close, using the
Trigger Input
tab
(Figure 4.38) for the test being run. The default configuration for this tab
is to have a
Trigger Source
of
None(Internal)
. It causes the external trigger
inputs to be inactive and causes the TDR9000 firmware to initiate a test
in response to input from the safety switch.
N
OTE
Selecting None(Internal) requires the use of the safety cable. Selecting
either external trigger input type requires the use of the safety bypass
flag.
Figure 4.38 Trigger Input Tab
The following parameters are configured on the Trigger Input tab:
Trigger Source
This picklist is used to select either:
•
AUX Contact
trigger
This is when the trigger comes through one
of the Auxiliary contacts of the Event module.
This selection does not automatically activate
the auxiliary contact channel for recording. The
channel must be activated via the Test Plan tab
also.
or
Содержание TDR9000
Страница 26: ...xxii 72A 1898 Rev A 11 01 Monday November 26 2001 1 09 pm...
Страница 50: ...Step 5 Saving Test Results 1 24 72A 1898 Rev A 11 01 Monday November 26 2001 1 09 pm...
Страница 98: ...3 28 72A 1898 Rev A 11 01 Monday November 26 2001 1 09 pm...
Страница 200: ...Disconnecting After the Test 4 102 72A 1898 Rev A 11 01 Monday November 26 2001 1 09 pm...
Страница 362: ...B 34 72A 1898 Rev A 11 01 Monday November 26 2001 1 09 pm...
Страница 388: ...D 6 72A 1898 Rev A 11 01 Monday November 26 2001 1 09 pm...
Страница 392: ...E 4 72A 1898 Rev A 11 01 Monday November 26 2001 1 09 pm...
Страница 394: ...F 2 72A 1898 Rev A 11 01 Monday November 26 2001 1 09 pm...
Страница 442: ...I 32 72A 1898 Rev A 11 01 Monday November 26 2001 1 09 pm...
Страница 454: ...I 12 72A 1898 Rev A 11 01 Monday November 26 2001 1 09 pm...