7

before using equipment. Pay attention to the ignition of endogenous gases.
•  The user should be exposed to the equipment noise during a maximum period of 8 hours/day.
•  Due to the reduced size of the equipment, some safety items could not be inserted, such as: means 
of protection against non-authorized use, control switch and emergency stop for the laser. Thus, the 
user shall keep the equipment in a safe and protected place.

EQUIPMENT SAFETY ITEMS

Laser Beam

The equipment uses a low intensity laser beam (red laser) to indicate the infrared laser irradiation 
point of the laser therapy handpiece.

As the laser beam passes through the same delivery system as the working beam, it offers a good 
way to check the integrity of the delivery system. If the target beam is not present in the distal end 
of the delivery system, if its integrity is reduced or if it seems diffuse, this is a possible indication that 
the delivery system is damaged or not working properly.

Alerts

For operator’s safety, during emission of laser light, the equipment sounds an alarm. 

Spacer

A spacer is sent to be inserted into the tip of the optical fiber, so as to have a contact application, 
i.e., the spacer is an applied part of the product. To replace the spacer, simply pull out the one that 
has already been used, clean the equipment and the new spacer, according to item “CLEANING/
DISINFECTING”, and insert it into the end of the optical fiber.

Handpiece Led

When the led of the handpiece is blinking fast, the laser has been activated.

LIST OF COMPONENTS

Therapy Plus is composed by:

PIECES:

Handpiece

Handpiece support

Power cable 

Содержание Therapy Plus

Страница 1: ...User Manual Therapy Plus ENG...

Страница 2: ...IN n 4 2012 established by ANVISA provides the documents for download at the following electronic address www dmcgroup com br in the USE INSTRUCTIONS page Check the revision of the User Manual and ANV...

Страница 3: ...diabetic ulcers cutaneous ulcers of different etiologies venous stasis ulcer skin scores vascular microcurrent diabetes post surgery contact dermatitis burns pressure ulcer diabetic ulcer and varicos...

Страница 4: ...lure Hematologic neoplasms tumors of hematopoietic and lymphoid tissues Uncontrolled glaucoma Pregnant patients Do not apply the laser on tattoos Aesthetic phototherapy On pregnant uterus Neoplasm in...

Страница 5: ...nt Output current Continuous current Input power 25 VA Working mode Continuous Voltage frequency 50 60 Hz Protection against shock Internally Powered Protection degree regarding water and solid object...

Страница 6: ...ver direct it for any person except he she is on treatment Bright surfaces may reflect laser light towards the eyes Never radiate tumor processes directly the laser can stimulate them Never radiate an...

Страница 7: ...heck the integrity of the delivery system If the target beam is not present in the distal end of the delivery system if its integrity is reduced or if it seems diffuse this is a possible indication th...

Страница 8: ...r transportation Additional parts or accessories may be purchased from the DMC through the following codes Code Description 010130131 Power cable 010990317 Power Supply 050020001 Safety glasses 050020...

Страница 9: ...is blinking fast the laser is active and when it is slow the battery is charging C Left button it selects the red or infrared laser D Right button it selects the energy in Joules and the Ilib techniqu...

Страница 10: ...ld be used in dental medical or physiotherapy offices and it should be posi tioned over a bench and connected to power outlet according illustrated below 1M The spacer inserted in the handpiece must b...

Страница 11: ...figure below Reinstall the lid of the battery compartment Before switching on the equipment for the first time it is necessary to charge the battery There are two ways to charge the battery connectin...

Страница 12: ...ain starting from the point of interruption press the switch button or press the right button to start a new application cycle At the end of the applica tioncycle thetimecounterisrestartedautomaticall...

Страница 13: ...INSULATION OF ELECTRICAL NETWORK In case of emergency or to clean the equipment power cable plug must be disconnected from outlet CLEANING DISINFECTION Before being cleaned the equipment should be di...

Страница 14: ...iece shall not be broken The optical fiber shall not be damaged Do not try to disassemble the handpiece or its support In case of problem in the equipment operation contact the DMC technical assistanc...

Страница 15: ...e risks the customer should discard the equipment according to the local legislation The Therapy Plus battery contains lithium and shall not be disposed in common waste LASER SUPPLY SYSTEM The equipme...

Страница 16: ...dA d E Maximum Power 0 25 W Area 1 2 x 10 7 m2 E 20 x 105 W m2 CALCULATION OF CORNEA MPE Applied Standard 60825 1 2014 Table A 2 Wavelength 808 nm 2 25 0 2 6 4 18 m W T C C EMPE t T2 C6 66 7 C4 1 64 T...

Страница 17: ...usto mer or user shall ensure it will be used in such environment Immunity test Compliance Electromagnetic Environment Standard RF emissions ABNT NBR IEC CISPR 11 Group 1 Compliance Therapy Plus uses...

Страница 18: ...ge IEC 61000 4 5 2 kV line to ground 1 kV line to line Compliance Quality of energy supply should be like a hospital or typical commercial environment Voltage drop short interruptions and vol tage var...

Страница 19: ...lower than compliance level of each frequency rangeb It may clash the equipment marked with the symbol Note 1 In 80 MHz and 800 MHz the highest frequency range is applied Note 2 These standards can t...

Страница 20: ...rding to transmitter frequency m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 37 0 37 0 74 1 1 2 1 2 2 3 10 3 7 3 7 7 4 100 12 12 23 For transmitters with maximum n...

Страница 21: ...equipment which will be connected to the direct current power supply Continuous current I P 20 Protected against 12 5 mm size solid unknown objects and not protected against water Manufacture date Man...

Страница 22: ...the equipment is repaired out of warranty period it can be only extended to exchanged components E The most common defects are physical shocks applied to the equipment and in these cases the warranty...

Страница 23: ...UIPAMENTOS LTDA Rua Sebasti o de Moraes 831 Jardim Alvorada S o Carlos SP CEP 13562 030 CNPJ 02 827 605 0001 86 ANVISA Registration 80030810157 Tech Responsible Renaldo Massini Jr CREA 0601706815 Tech...

Страница 24: ...For more information about our technology and how to use it visit www nupen com br Technical Assistance DMC IMP E EXP DE EQUIP LTDA 55 16 2107 2323 www dmcgroup com br...