The ARTROMOT®-E2/-E2 compact is
suitable for re-use if the following points
are observed.
6.1 Care/ Re-use
m
Warning!
Shock hazard – Unplug the device from
the power line before cleaning.
Shock hazard, equipment damage –
Liquids must not enter the device or the
programming unit.
• The ARTROMOT®-E2/-E2 compact must
be
disinfected by wiping down
with a
disinfectant. Thus, it complies with the
special hygiene standards for medical
technical equipment.
•
The enclosure
and removable
armrest
assemblies
can be cleaned with
com-
monly used disinfectants
and
mild
household detergents
.
• Only use a
damp cloth
to wipe the CPM
device down.
m
Warning!
Patient hazard –
patient contamination
− Before using the device on another
patient, be sure to clean and disinfect
it according to the instructions given
here.
Note!
The manufacturer recommends using
only special disinfectants approved for
medical devices and with the character-
istics specified below under “Caution”.
Disinfection with the following disin-
fectants and in compliance with the
instructions for use has been evaluated
and approved by the manufacturer in
the risk assessment:
– DESCOSEPT AF lemon
(Art.-Nr: 00-311L-xxx)
BAuA Reg.No.: N-55153, -0482
Manufacturer: Dr. Schumacher GmbH
Contact time: 2 minutes minimum
m
Caution!
Equipment damage –
− The plastic material used is not
resistant to mineral acids, formic acid,
phenols, cresols, oxidants and strong
organic or inorganic acids with a pH
value below 4.
– Use only clear disinfectants to prevent
discoloration of the device.
– Do not expose the CPM device to
strong ultraviolet radiation (sunlight)
and fire.