background image

dIXEL

Operating and Instructions Manual

1592006100

1592006100  XB570L GB r1.0 04.11.2004.doc

XB570L

3/6

rS2 (

-50

÷

50°C; 1°C/1°F) 

Room probe Set point 

for the second phase: it

prevents temperature from reaching a too low value during the second
phase.

Pd2 

OFF

÷

4.0h;

 res. 

10 min

Maximum time for second phase.

iS3 (

-50

÷

50°C; 1°C/1°F) 

Insert Probe Set point

 to stop the third (and last)

phase: when the temperature measured by the third probe reaches this
value the whole cycle is ended.

rS3  (

-50

÷

50°C; 1°C/1°F)

 Room probe Set point

 for the third (and last) phase:

it prevents temperature from reaching a too low value during the third
(and last) phase.

Pd3 (

OFF

÷

4.0h; 10 min) 

Maximum time for the third phase.

dbH (yes / no) defrost before the hold phase
HdS (-50÷50 - OFF; 1 °C / 1°F) Set point of the holding phase

. With “OFF”

the hold phase is disabled.

IMPORTANT NOTE: 

If the duration time of a phase  is set at the OFF value,

the corresponding phase is disabled. E.g. If  

Pd3

= 0FF the third phase of the

cycle is not active.

14.2  . HOW TO USE THE INSERT PROBE.

By means the insert probe, the internal temperature of products can be
checked. This measure is used to end the various phase of the cycle. A
special internal function detect if the inset probe is not used, in this case the
cycle is made by time

14.3  EXAMPLE OF A BLAST CHILLER CYCLE.

The following drawing explains how a Blast Chiller cycle can be done.

14.3.1  First phase: “Hard chill”.

It is normally used to fast chill hot foods. E.g. from 80°C / 170°F to 20°C /
70°F
During “

Hard Chill

”, both compressor and fan are always on until the

rS1

 temperature is reached. At this point compressor is turned on end

off so as to keep the temperature of the room at the 

rS1

 value. “Hard

Chill” ends when the temperature measured by the third probe reaches
the 

iS1

 value. Normally 

rS1

 is set at few degrees below zero.

14.3.2 Second 

phase: “Soft chill”.

The 

Soft Chill

 starts when the Hard Chill ends. It is used to prevent thin

layer of ice from forming on the product. The Soft Chill lasts until the
temperature measured by the third probe reaches the set point 

iS2

(usually 4 or 5°C).
During Soft Chill the temperature of the room is regulated by the ambient
probe with the set point 

rS2

 (normally at 0 or 1 °C / 32 or 34°F). When

the box temperature reaches the 

rS2

 value compressor is turned on end

off so as to keep the temperature of the box at this value.

14.3.3  Third phase: “Freezing cycle”.

Freezing Cycle: used to fast freeze foods.
The Freezing Cycle starts when the Soft Chill ends. During the “Freezing
Cycle” both compressor and fan are always on until the 

rS3

 temperature

is reached. At this point compressor and fans are turned on end off so as
to keep the temperature of the room at the 

rS3

 value (normally some

degrees below 

iS3)

. Freezing Cycle ends when the temperature

measured by the third probe reaches the 

iS3

 value (normally -18°C /

0°F), in any case it ends when the maximum time 

Pd1 + Pd2 + Pd3

 has

expired.

14.3.4  End of the Blast Chill cycle and starting of the Hold Mode.

When the Blast Chiller cycle ends an alarm signal is generated: buzzer
and alarm relay is turned ON, the display shows the message “End”
alternating with the room temperature.

The alarm automatically stops after 30min or by pressing any key.
At the end of the Blast Chiller cycle the controller can start the “Hold
mode” keeping the room temperature at the value set in HdS parameter.
If HdS = OFF, the machine is turned OFF.

NOTE1:

 with 

dbH = yES

 a defrost is done before the holding phase.

NOTE2: 

If the end cycle temperature IS3 is not reached by the maximum

time Pd1+Pd2+Pd3 the instrument keep on working, but the alarm message

“OCF” 

is given.

15.  Function And Parameter Programming Menu

15.1 FUNCTION 

MENU

Includes all the main functions controlled by the instrument.

15.2 ACCESS 

PROCEDURE:

1. Hold 

the 

SET

 and 

DOWN

 keys pressed for few seconds, until the label of

the first function will be displayed.

2. The 

UP

 and 

DOWN

 keys are used to cycle backwards or forward in the

menu.

3.  By pressing the 

SET

 key the currently displayed function is enabled.

4.  To go back to the previously Menu press the 

SET

 and 

DOWN

 keys

together

15.3  LIST OF FUNCTIONS

15.3.1  “Prb”: probe display

It permits the user to display the temperatures measured by the probes.

 

rPr: 

displays for 10s the temperature measured by the room probe.

 

EPr: 

displays for 10s the temperature measured by the evaporator

probe.

 

iPr: 

displays for 10s the temperature measured by the insert probe.

NOTE1

 If a key is pressed when a temperature is displayed, the

instrument go back to show the label of the probe (rPr, EPr, iPr).

NOTE2: 

If the selected probe is broken, the follow alarm labels will be

displayed: 

rPF 

for the room probe, 

EPF

 for the evaporator probe, 

iPF

 for

the insert probe.

15.3.2  “Pr1”: user parameters

 

“Pr1” 

includes all user accessible parameters. If no parameter would be

present in the menu, the label “Pr2” will be displayed.

15.3.3  “Pr2”: installer parameters

 

Pr2

”: includes all the instrument’s parameters (at installer level). It can be

accessed through a security code. Hence it is possible to modify all
parameters and add or remove parameters from “Pr1” (user level) by
pressing 

“SET”

 + 

“UP”

. When a parameter is enabled at user level,

LED 

 (Alarm LED) is on.

15.3.4  “FCy”: cycle management

The FCy menu contains the Cy1, Cy2, Cy3, Cy4 submenu. Each
submenu contains all the configurable parameter of the cycle as shown
in the following drawing:

Cy1

Fcy

Cy2

Cy3

Cy4

rS1

Pd1

dbH

iS2

rS2

Pd2

iS3

rS3

Pd3

iS1

dbC

HdS

rS1

Pd1

dbH

iS2

rS2

Pd2

iS3

rS3

Pd3

iS1

dbC

HdS

rS1

Pd1

dbH

iS2

rS2

Pd2

iS3

rS3

Pd3

iS1

dbC

HdS

rS1

Pd1

dbH

iS2

rS2

Pd2

iS3

rS3

Pd3

iS1

dbC

HdS

15.3.5  “Sto”: cycle length

The 

Sto

 menu records the real length of a cycle and the real length of

each single phase. It contains the following sub-menu:

tCy

: cycle duration, 

tP1

: first phase duration; 

tP2

: second phase

duration; 

tP3

: third phase duration.

Содержание XB570L

Страница 1: ...tures The series XB has been created for fast chilling or freezing goods according to international food safety standards There are FOUR types of cycles The CYCLES Cy1 Cy2 Cy3 Cy4 are pre set according to the most common cycles used in food safety applications the user can select one of them according to his own requirements and modify it as he wants Any cycle can be manually terminated before the...

Страница 2: ...ys C1 C2 C3 C4 the related led will be lighted and the cycle will be selected NOTE to pass from a cycle to another one simply push the related key HOLD PHASE To select the old phase deselect the enabled cycle pushing the related key ES Cycle 1 is selected its led is oon push the C1 key the led will be turn off and the hold phase will be enabled 2 Cycles are pre set with the following values 1 Cy1 ...

Страница 3: ...eaches the iS3 value normally 18 C 0 F in any case it ends when the maximum time Pd1 Pd2 Pd3 has expired 14 3 4 End of the Blast Chill cycle and starting of the Hold Mode When the Blast Chiller cycle ends an alarm signal is generated buzzer and alarm relay is turned ON the display shows the message End alternating with the room temperature The alarm automatically stops after 30min or by pressing a...

Страница 4: ...sor ON time with faulty probe 0 255 min time during which the compressor is active in case of faulty thermostat probe With COn 0 compressor is always OFF COF Compressor OFF time with faulty probe 0 255 min time during which the compressor is off in case of faulty thermostat probe With COF 0 compressor is always active diC Digital input operating mode EAL bAL dor EAL external alarm bAL serious alar...

Страница 5: ...the DOWN key or by closing the door 19 1 3 OAC UL UV light for a sterilisation cycle It is used for a sterilisation cycle of the machine It can be enabled ONLY WHEN THE CONTROLLER IS NOT WORKING or when a manual defrost is running The ultra violet light stays on for the OAt time It can be switched off before the OAt time by pressing the DOWN key NOTE with OAt 0 it can be switched off only by press...

Страница 6: ... rS2 30 C 22 F Pd3 OFF rS1 30 C 22 F Pd2 OFF dbH yes Pd1 4 0 iS3 18 C 0 F HdS 18 C 0 F Cy4 direct fast freezing without holding dbC no iS2 18 C 0 F rS3 30 C 22 F iS1 18 C 0 F rS2 30 C 22 F Pd3 OFF rS1 30 C 22 F Pd2 OFF dbH no Pd1 4 0 iS3 18 C 0 F HdS OFF 24 Standard Values of the parameters Lab Description XB570L C F Regulation Hy differential 2 4 AC Anti short cycle delay 1 1c2 Second compressor ...

Отзывы: