Display modes
6. Display modes
Upon activation, the analyzer is normally in the O2/He measuring mode, in one of the two
display formats.
The
basic format
shows the oxygen and helium content in the mixture in percent in separate
lines and, for information, also the temperature at which the gas was measured.
After the measuring begins, the helium concentration may appear in square brackets for
several seconds in the basic format. This means that the result is interim, with lower accuracy.
When the brackets disappear, the accuracy of the result is final.
In case of a measuring error, for instance at a pneumatic shock in the inlet hose, the value
of the measured helium concentration disappears for the time of duration of the error state.
The
well-arranged format
shows the composition in a way used by divers to call the mixtures:
— Air: air with the oxygen content between 20.5 % and 21.5 %.
— EAN, for instance EAN 36, is air enriched with oxygen, in this case up to 36 % oxygen.
— TMX, for instance TMX 18/45, is trimix or the mixture of air, oxygen and helium, in this
particular case containing 18 % oxygen and 45 % helium.
— Heliox, for instance Heliox 16 % O2, is a mixture of oxygen and helium.
— Foul Air, for instance Foul Air (10 % O2), is foul or vitiated air with a reduced oxygen content.
It cannot be mixed with gases designed for preparation of breathing mixes commonly avail-
able to the diver (air, oxygen, helium); therefore, its presence indicates a problem. This may
include ongoing corrosion inside the bottle or contamination of the mix with argon, carbon
dioxide or another gas. Such air cannot be used for breathing under any circumstances.
In addition to the mix composition, this well-arranged format shows approximate values of the
maximum operating depth (MOD) and an equivalent narcotic depth (END).
MOD indicates the depth to which the diver can dive for a short time under otherwise ideal
conditions. The limit of partial oxygen pressure equal to 1.6 kPa is used in the calculation,
and seawater and the static pressure at sea level are taken into account. This reading is for
orientation only, and for an actual dive it must be calculated using the data of the specific dive.
The following formula is used for the calculation:
2
max
O
0
P
R
P
MOD
g
ρ
−
=
kde
P
max
is the maximum allowable partial oxygen pressure, chosen as 160,000 Pa
P
0
is the static pressure at sea level, 101,325 Pa
R
O2
relative molar concentration of oxygen in the interval 0 to 1
ρ
is the seawater density, 1028 kg m
−3
g
is the standard acceleration of gravity, 9.80665 m s
−2