Page 9
Step 5 Set up Menu
At the main display long press the
tick key to see the Set up
menu. This has 6 options:
5.1 Select units
To Select the Units of
measurement, long press the
tick key to get to the setup
menu, then press the
tick key
again and the units menu will be displayed. The units
available are mT (milliTesla), T (Tesla), G (Gauss), kA/m
(kiloamps metre), Oe (Oersteds) and kA (kiloamps) at
distance. To select a new system of units, short press the
up or
down keys until the required units are indicated.
Then short press
tick to set the value of the magnetic
field in the selected units. In selecting kA (kiloamps) at
distance, you will be prompted to enter the distance from
the conductor at which the measurement will be made.
5.2 Average
This allows the number of
samples to be selected for
averaging. This is normally set to
64, but any number in the set of
[1, 4, 16, 64, 256] can be selected. The choice of this
value controls the time response of the instrument in
average mode. A small number will result in a fast
response but with greater reading uncertainty. A large
number will make the display update sluggish, but will
improve noise by corresponding factors of [1, 2, 4, 8, 16].
Operation - Sep up