background image

Do not tighten the SomnuSeal straps. 
 

9.  Now you are ready for sleep.

 

 

 

 

 

 

 

 

 

 

On waking up 
1.  Remove the Straps from your head and the Body 

of the mask from your mouth. 

2.  Turn the PAP machine and humidifier off. 
3.  Detach the Strap Connector and the Straps from 

the main Body of the mask. 

4.  Detach the SomnuSeal Air Vent from the Strap 

Connector. 

5.  Thoroughly clean the main Body of the mask 

according to the instructions in the Maintenance 
section of this document. 

6.  The Strap Connector and Straps do not need to be 

cleaned every day and should not get wet while 
you are cleaning the main Body of the mask.  
  
SomnuSeal Maintenance 

 
 

  Store the mask and its accessories in a cool dry 

place  
(5-50 Celsius = 40-120 Fahrenheit). 

  Do not expose the mask to direct sunlight for a 

prolonged period of time. 

  Replace and discard the mask if it shows any 

mechanical deterioration such as deformation 
and/or color change. 

 
 
 

 
 
For home users: 
Every Day: 
1.  Detach the Strap Connector (with the Straps and 

the SomnuSeal Air Vent attached to it) from the 
main Body of the mask. 
Detach the SomnuSeal Air Vent from the strap 
connector.  

2.  Thoroughly wash the main Body of the mask 

and the SomnuSeal air vent with warm water 
and soap. Make sure you also clean the inner 
part of the main Body and the SomnuSeal Air 
Vent. Please refrain from using sharp objects or 
finger nails during the cleaning process. 
Rinse with lukewarm water and place to dry out 
of direct sunlight. 
Every 7 days: 

3.  Hands wash the straps with soap dissolved in 

lukewarm water and place to dry out of direct 
sunlight. 
 

 
For institutional users: 
1.  Detach the Strap Connector (with the Straps and 

the SomnuSeal Air Vent attached to it) from the 
main Body of the mask. 
Detach the SomnuSeal Air Vent from the strap 
connector.  

2.  Immerse the main Body of the mask and the 

SomnuSeal air vent completely, filling all lumens 
and eliminating air pockets, in a CIDEX OPA 
Solution for a minimum of 12 minutes at 20°C or 
higher. 

3.  After removing the Body of the mask and the 

SomnuSeal air vent from the CIDEX OPA 
Solution, thoroughly rinse it by immersing it 
completely in a large volume (e.g., 8 liters) of 
water. 

4.  Keep the mask’s Body and the SomnuSeal air 

vent completely immersed for at least one (1) 
minute. 
Manually flush all lumens with large volumes of 
rinse water (at least 100ml). 

5.  Remove the Body of the mask and the 

SomnuSeal air vent and discard the rinse water.  

6.  Repeat the procedure two (2) more times, for a 

total of three (3) rinses, with large volumes of 
fresh water. 

7.  Hand wash the straps with antiseptic soap 

dissolved in lukewarm water and place to dry out 
of direct sunlight. 

 
 
 

  The SomnuSeal mask and the SomnuSeal air 

vent must be disinfected before treating 
another patient. 

  You must perform three (3) separate, large 

volume water immersion rinses,  
to remove all CIDEX OPA Solution residues, as 
they may cause serious side effects.  

  Always use fresh volumes of water for each 

rinse. Do not reuse the water for any purpose. 
Use sterile water unless potable water is 
acceptable. 

 

 