background image

DIRECTHEALTHCAREGROUP.COM

USER MANUAL

9

Castors/brakes

Castors roll easily/tread not damaged.
The castors are tightly fastened to the lower frame.
Castor brakes functional on all castors.
Handbrake functional.

Clean or replace castors (NOTE: Castors are 

always fixed to the lower frame with thread locker 

or lock nuts. The castors are sealed and we do not 

recommend their dismantling; instead, replace the 

whole castor assembly.

Tighten the castor bolt and fit locking nut or use 

thread locker (depends on model).
Replace castor assembly.

Adjust the brake or fit new castors.

Harness 

Belts, seams and Velcro fasteners are whole/intact.
The walker lockplate (green) is functional.

Holder, belts

Holder, belts not bent (see Figure s3).
Harness and attachment belts must be replaced
every three years.

Replace harness. Available in three sizes (S, M, & L)
Replace lockplate

A bent holder is a sign that the product has been 
overloaded. The frame must be replaced before 
further use.
The action is then ”Check harness labeling. If the 
harness production date is older than 3 years, the 
harness and belts must be scrapped.

12. General maintenance advice

•  Before use, check that the holder and belts are not bent (see Figure s3).

•  Before use, check that the harness is not more than 3 years old (see labeling).

• 

Wipe the walker and its armrests clean as necessary; use soap solution/surface disinfectant.

•  The product can be washed in washer/machine or similar since electronic components are sealed to IP65 

classification. Please note that the product is not made out of stainless material – frequent use of water/

detergents may therefore shorten the life of the product.

• 

Replace damaged parts immediately; must be done by experienced personnel.

•  The product is only intended for indoor use.

13. Technical specifications

Dimensions apply to walkers with 125 mm castors.

Part No.

Product 
width

Max user 
weight

Length

Height

Weight

Armrest/ 
Internal 
dimension

56-316

77cm

150kg

89cm

96-136cm

25kg

34-65cm

Содержание Bure Rise & Go

Страница 1: ...DIRECT HEALTH C ARE GR OUP COM Bure Rise Go User Manual ...

Страница 2: ...BURE RISE GO DIRECTHEALTHCAREGROUP COM 2 ...

Страница 3: ... everyday life easier thanks to its smart power rise function The Bure Rise Go can be used by anyone who regularly uses a walker The user can remain active while the power rise function is in use The product consists of a walker and its associated harness and straps The harness is available in different sizes S M L XL and is split at the rear to make remaining in the harness possible while using t...

Страница 4: ...h their feet on the floor or platform Adjust the shin supports to fit the user s shins remember to avoid pressure on the patellae and patellar tendons Refer to page 5 for shin support adjustment 3 Thread both attachment belts through the walker s frame and the green locking devices Make sure the belts are threaded completely through the locking devices NB Check especially that the belts are undama...

Страница 5: ...applying their weight to the stirrup and maintaining eye contact with the patient The other carer may stand next to the patient 8 In most cases the patient will not achieve a fully upright standing position encourage the patient to achieve a final upright posture independently or assist the patient by tightening the adjustable harness straps 9 Lower the walker to its individually adjusted height a...

Страница 6: ...one foot firmly on the extended stirrup in order to prevent the walker from tipping backwards 5 Left and right adjustment of armrests The armrests can be adjusted sideways for optimal support the whole arm can be moved in or out and locked with the knob 6 Locking unlocking castors Depress the foot brake to set it and lock the castor To unlock press down on the protruding part of the brake 7 Adjust...

Страница 7: ... the battery is switched off by means of the ON OFF switch This eliminates the risk of disabling the battery as the control box always uses a small quantity of electricity even when the walker is not in use 10 LED on controlbox and battery Control box green light when power is available from battery mains socket Battery The LED flashes green during charging ON OFF To use the product press ON The g...

Страница 8: ...t loose Fit new cable this must always be connected to the control box IP65 classification Connect replace with new remote control as required Refer to the wiring diagram in the Technical manual Tighten replace with new fasteners as required Height adjustment Raising and lowering are functional The walker feels stable at maximum settings There is no play between the vertical frame and bottom frame...

Страница 9: ...t the product has been overloaded The frame must be replaced before further use The action is then Check harness labeling If the harness production date is older than 3 years the harness and belts must be scrapped 12 General maintenance advice Before use check that the holder and belts are not bent see Figure s3 Before use check that the harness is not more than 3 years old see labeling Wipe the w...

Страница 10: ...BURE RISE GO DIRECTHEALTHCAREGROUP COM 10 ...

Страница 11: ...DIRECTHEALTHCAREGROUP COM USER MANUAL 11 ...

Страница 12: ...THEALTHCARE GR OUP COM Direct Healthcare Group Withey Court Western Industrial Estate Caerphilly CF83 1 BF T 0800 043 0881 F 0845 459 9832 E info directhealthcaregroup com LIT 00126 Issue 1 Date January 2020 ...

Отзывы: