UltiMate
®
3000 Viper™ Capillary Kits for Intelligent LC (LCi) System Solutions
Quick Installation Guides
Page 20
Rev. 02
January 2010
©
2010 Dionex
Table 6:
Detailed description of capillary connections of the automated application switching setup
Description of Connections
Capillary Dimensions (ID x L)
PumpLeft outlet – TCC 10-port valve right, port 2
0.18 mm/0.007” x 850 mm
PumpRight outlet – TCC 10-port valve right, port 4
0.18 mm/0.007” x 750 mm
TCC 10-port valve right, port 1 – Column 1 inlet
RS: 0.13 mm/0.005” x 350 mm
SD: 0.18 mm/0.007” x 350 mm
TCC 10-port valve right, port 5 – Column 2 inlet
RS: 0.13 mm/0.005” x 350 mm
SD: 0.18 mm//0.007” x 350 mm
WPS sampler valve, port 5 – TCC 10-port valve right, port 3
RS: L 0.13 mm//0.005” x 350 mm
SD: 0.18 mm//0.007” x 350 mm
TCC 10-port valve right, port 10 – TCC 6-port valve left, port 6
RS: 0.13 mm//0.005” x 650 mm
SD: 0.18 mm//0.007” x 650 mm
TCC 10-port valve right, port 6 – TCC 6-port valve left, port 2
RS: 0.13 mm//0.005” x 650 mm
SD: 0.18 mm/0.007” x 650 mm
WPS sampler valve, port 4 – TCC 6-port valve left,
port 1
RS: 0.13 mm//0.005” x 750 mm
SD: 0.18 mm/0.007” x 750 mm
TCC 10-port valve right, port 7 – Waste
0.25 mm/0.010” x 1,000 mm
(PEEK)
TCC 10-port valve right, port 9 – Waste
0.25 mm//0.010” x 1,000 mm
(PEEK)
Column 1 outlet – TCC 6-port valve left, port 5
RS: 0.13 mm/0.005” x 250 mm
SD: 0.18 mm/0.007” x 250 mm
Column 2 outlet – TCC 6-port valve left, port 3
RS: 0.13 mm/0.005” x 250 mm
SD: 0.18 mm/0.007” x 250 mm
TCC 6-port valve left, port 4 – Detector inlet
RS: 0.13 mm/0.005” x 350 mm
SD: 0.18 mm/0.007” x 350 mm
Note:
In case of a dual-stack setup, the TCC 10-port valve
right
turns into a
TCC 10-port valve
left
and the TCC 6-port valve
left
turns into a TCC 6-port
valve
right
. Both stacks should have a distance of about 7 cm (2.8 inches). The
capillary
connection between valve right and detector
has to be
extended
from 350 mm to
550 mm
. Connecting a post-column cooler with a stackable
Dionex detector does not work without additional capillaries when setting up
the system in the U-configuration (