IonPac TCC-LP1, TCC-ULP1, and TCC-XLP1 
Product Manual 

Page 15 of 17 

 

 

 

 

 

 

Doc. No. 034973-06 

©2005 DIONEX 

September 2005 

 

 

5.

 

TROUBLESHOOTING GUIDE 

The purpose of the Troubleshooting Guide is to help you solve operating problems that may arise while using the TCC-LP1, 
TCC-ULP1, and TCC-XLP1 columns.  For more information on problems that originate with the Ion Chromatograph, refer to 
the Troubleshooting Guide in the appropriate operator’s manual.  If you cannot solve the problem on your own, call your 
nearest Dionex Regional Office (see, “Dionex Worldwide Offices” on the Dionex Reference Library CD-ROM). 
 

5.1.

 

High Backpressure from a Contaminated Inlet Bed Support 

If the TCC-LP1, TCC-ULP1, or TCC-XLP1 column displays high backpressure, the bed support in the column inlet may be 
contaminated.  A contaminated bed support may also lead to loss of peak asymmetry. 
 

WARNING

 

For proper operation in the SP10 Auto Neutralizer, the TCC columns should generate no more than 
70 psi of backpressure at a flow rate of 0.5 mL/min.

WARNING

If any of the column packing becomes lodged between the end of the column and the bed support 
washer assembly, no amount of tightening will seal the column.  Be sure that the washer and the end 
of the column are clean before screwing the end fitting back on to the column.

 
Follow the instructions below to change the bed support assembly using one of the two spare bed support assemblies 
included in the ship kit provided with the column. 
 

A.

 

Disconnect the column from the system. 

 

B.

 

Using two open-end wrenches, carefully unscrew the inlet (top) column end fitting. 

 

C.

 

Turn the end fitting over and tap it against a bench top or other hard, flat surface to remove the bed support and 
seal assembly.  If the bed support must be pried out of the end fitting, use a sharp pointed object such as a pair 
of tweezers, but be careful that you DO NOT SCRATCH THE WALLS OF THE END FITTING.  Discard the 
old assembly. 

 

D.

 

Place a new bed support assembly in the end fitting.  Before assembling, clean the column body threads and the 
end fitting threads of any resin particles.  If any resin remains on the threads of either the column body or the 
end fitting, the column may leak regardless of how tight the end fitting is turned onto the column body.  Use the 
end of the column to carefully start the bed support assembly into the end fitting. 

 

E.

 

Screw the end fitting back onto the column.  Tighten it finger tight and then, using two open-end wrenches, 
tighten it an additional 1/4 turn (25 in x lb).  Tighten further only if leaks are observed. 

 

F.

 

Reconnect the column to the system and resume operation. 

 
 

Column 

TCC-LP1 (P/N) 

TCC-ULP1 (P/N) 

TCC-XLP1 (P/N) 

Bed Support Assembly 

042955 

042955 

063688 

End Fitting (10-32 Ferrule 

Type) 

052809 052809  048298 

 
 
 

Содержание IonPac TCC Series

Страница 1: ...IonPac TCC...

Страница 2: ...ptember 2005 PRODUCT MANUAL FOR TRACE CATION CONCENTRATOR TCC TCC LP1 Column Low Pressure 4 x 35 mm P N 063079 TCC ULP1 Column Ultra Low Pressure 5 x 23 mm P N 063783 TCC XLP1 Column Extremely Low Pre...

Страница 3: ...acity 10 4 EXAMPLE APPLICATIONS 13 4 1 Concentration in the Manual Mode 13 4 2 Concentration in the Manual Mode 14 5 TROUBLESHOOTING GUIDE 15 5 1 High Backpressure from a Contaminated Inlet Bed Suppor...

Страница 4: ...throughout this document to emphasize important points The symbols are shown below along with the purpose of the information SAFETY Safety information can help prevent bodily harm WARNING Warning inf...

Страница 5: ...CC ULP1 is less than 45 psi 0 31 MPa at 1 mL min and the TCC XLP1 is less than 30 psi 0 21 MPa at 1 mL min The large resin particle size 20 m in the TCC LP1 TCC ULP1 and TCC XLP1 makes it possible to...

Страница 6: ...ional Group Hydro phobicity TCC LP1 4 mm 20 55 260 eq col Carboxylic Acid Very low TCC ULP1 5 mm 20 55 260 eq col Carboxylic Acid Very low TCC XLP1 6 mm 20 55 260 eq col Carboxylic Acid Very low TABLE...

Страница 7: ...Column Description The IonPac Trace Cation Concentrator Low Pressure TCC LP1 4x35 mm and IonPac Trace Cation Concentrator Ultra Low Pressure TCC ULP1 5x23 mm columns consist of the following component...

Страница 8: ...ugh to ensure good quantification WARNING The flow direction during the concentration step is critical In order to ensure optimal system performance it is recommended that concentration always be perf...

Страница 9: ...IONEX September 2005 Figure 3 shows the configuration for sample loading using a Rheodyne valve FIGURE 3 Configuration for Determining Trace Levels of Cations Figure 4 shows the configuration for samp...

Страница 10: ...o 8 days Recommended storage vessels are Corning tissue culture flasks The following procedure should be used for storage of g L level samples A Rinse the polystyrene container and cap twice with deio...

Страница 11: ...he analytes on the ion exchange sites and result in their elution to waste during the loading process B Conversely qualitative analysis of ions with higher affinities for the resin in the presence of...

Страница 12: ...e rate at which the sample is loaded C The pH of the sample D The ionic strength of the sample E The amount and capacity of resin in the column The breakthrough volume is determined as follows A Prepa...

Страница 13: ...2005 DIONEX September 2005 FIGURE 6 Linearity Determinations for Concentrator Injection Time min that the simulated sample is pumped through the TCC LP1 TCC ULP1 or TCC XLP1 after the single ion injec...

Страница 14: ...thanesulfonic acid Eluent Flow Rate 1 0 mL min Injection Volume 1 0 mL Temperature 35 C SRS Suppressor Cation Self Regenerating Suppressor CSRS ULTRA AutoSuppression Recycle Mode MMS Regenerant 100 mN...

Страница 15: ...n Eluent 10 mM Methanesulfonic acid Eluent Flow Rate 1 0 mL min Injection Volume 1 0 mL SRS Suppressor Cation Self Regenerating Suppressor CSRS ULTRA AutoSuppression Recycle Mode or AES Suppressor Cat...

Страница 16: ...back on to the column Follow the instructions below to change the bed support assembly using one of the two spare bed support assemblies included in the ship kit provided with the column A Disconnect...

Страница 17: ...D Be sure that the eluents are made from chemicals with the recommended purity see Section 3 Operation E Be sure that the deionized water used to prepare the reagents has a specific resistance of 18...

Страница 18: ...he gradient pump Direct the effluent from the TCC LP1 TCC ULP1 or TCC XLP1 directly to a waste container C Set the flow rate to 1 mL min D Pump the 1 M HCl 0 1 M KCl solution through the column for 15...

Отзывы: