This error occurs if you attempt to use relay 4 as a standard relay in a
Chromeleon or Chromeleon Xpress program, although the relay has been set
to synchronize the gradient with the autosampler sample injection.
To troubleshoot:
Define the relay as a standard relay
or
specify a different relay for use in the
program.
If the DP/SP pressure exceeds 5 MPa (725 psi) during priming, this error
occurs and the priming process is aborted.
To troubleshoot:
1. Verify that the priming valve is open and then reprime the pump. (To open
the valve, turn the knob one-half turn counterclockwise.)
2. Check fittings for overtightening. Check tubing for crimping and replace
it as necessary. Be careful not to overtighten fittings.
This error may be caused by a blockage in the system or by a problem with
the column. (“Right-hand” refers to the bottom pump; “left-hand” refers to the
top pump in a dual-pump module.)
Relay 4 is configured for inject synchronization. Please
change pump configuration.
The maximum purge pressure was exceeded.
The pressure in the left-hand working cylinder exceeded the
safety limit.
-or-
The pressure in the right-hand working cylinder exceeded the
safety limit.
-or-
The system pressure exceeded the safety limit.
Содержание ICS-3000
Страница 1: ...ICS 3000 Ion Chromatography System Operator s Manual Document No 065031 Revision 04 January 2008...
Страница 32: ...ICS 3000 Ion Chromatography System 16 Doc 065031 04 1 08...
Страница 93: ...2 DC Description Doc 065031 04 1 08 77 Figure 2 42 DC Control Panel...
Страница 99: ...2 DC Description Doc 065031 04 1 08 83 Figure 2 46 Conductivity Detector Control Panel...
Страница 106: ...ICS 3000 Ion Chromatography System 90 Doc 065031 04 1 08...
Страница 138: ...ICS 3000 Ion Chromatography System 122 Doc 065031 04 1 08...
Страница 166: ...ICS 3000 Ion Chromatography System 150 Doc 065031 04 1 08...
Страница 174: ...ICS 3000 Ion Chromatography System 158 Doc 065031 04 1 08...
Страница 180: ...ICS 3000 Ion Chromatography System 164 Doc 065031 04 1 08...
Страница 218: ...ICS 3000 Ion Chromatography System 202 Doc 065031 04 1 08...
Страница 228: ...ICS 3000 Ion Chromatography System 212 Doc 065031 04 1 08...
Страница 272: ...ICS 3000 Ion Chromatography System 256 Doc 065031 04 1 08...
Страница 296: ...ICS 3000 Ion Chromatography System 280 Doc 065031 04 1 08...
Страница 354: ...ICS 3000 Ion Chromatography System 338 Doc 065031 04 1 08...
Страница 362: ...ICS 3000 Ion Chromatography System 346 Doc 065031 04 1 08...
Страница 368: ...ICS 3000 Ion Chromatography System 352 Doc 065031 04 1 08...
Страница 372: ...ICS 3000 Ion Chromatography System 356 Doc 065031 04 1 08...
Страница 374: ...ICS 3000 Ion Chromatography System 358 Doc 065031 04 1 08...