8 • Troubleshooting
Doc. 065031-04 1/08
187
with pressurized air or nitrogen. If the electrode is pitted, polish the
electrode (see
3. Check the value of the reference potential shift. See
Voltmeter to Determine Reference Potential Shift” on page 226
value fluctuates by more than 30 mV, electrode passivation may occur
because potentials that are too high are being applied. Recalibrate the
reference electrode (see
). If the problem persists, replace
Analog Output Option
•
Analog output range set too high
Select a more sensitive analog output range.
8.11 High Background
•
CR-TC is contaminated
Clean the trap column as instructed in the CR-TC manual. The manual is
included on the Dionex Reference Library CD-ROM (P/N 053891).
•
Wrong eluent
Check that you are using the correct eluent for your application. For ED
detectors, verify that the pH readout is correct for your eluent.
•
Background signal not offset from detector signal
Before injecting sample, allow the background signal to equilibrate, and then
press
Autozero
on the Control panel in Chromeleon or Chromeleon Xpress.
CD Only
•
Background not suppressed by suppressor
Verify that the suppressor is turned on and the current is set to the correct
value. Refer to the suppressor manual for additional troubleshooting
guidance. Suppressor manuals are included on the Dionex Reference Library
CD-ROM (P/N 053891).
•
Regenerant fails to suppress background
Use a higher regenerant flow rate. Refer to the suppressor manual for the
suggested regenerant flow rate for your application.
Содержание ICS-3000
Страница 1: ...ICS 3000 Ion Chromatography System Operator s Manual Document No 065031 Revision 04 January 2008...
Страница 32: ...ICS 3000 Ion Chromatography System 16 Doc 065031 04 1 08...
Страница 93: ...2 DC Description Doc 065031 04 1 08 77 Figure 2 42 DC Control Panel...
Страница 99: ...2 DC Description Doc 065031 04 1 08 83 Figure 2 46 Conductivity Detector Control Panel...
Страница 106: ...ICS 3000 Ion Chromatography System 90 Doc 065031 04 1 08...
Страница 138: ...ICS 3000 Ion Chromatography System 122 Doc 065031 04 1 08...
Страница 166: ...ICS 3000 Ion Chromatography System 150 Doc 065031 04 1 08...
Страница 174: ...ICS 3000 Ion Chromatography System 158 Doc 065031 04 1 08...
Страница 180: ...ICS 3000 Ion Chromatography System 164 Doc 065031 04 1 08...
Страница 218: ...ICS 3000 Ion Chromatography System 202 Doc 065031 04 1 08...
Страница 228: ...ICS 3000 Ion Chromatography System 212 Doc 065031 04 1 08...
Страница 272: ...ICS 3000 Ion Chromatography System 256 Doc 065031 04 1 08...
Страница 296: ...ICS 3000 Ion Chromatography System 280 Doc 065031 04 1 08...
Страница 354: ...ICS 3000 Ion Chromatography System 338 Doc 065031 04 1 08...
Страница 362: ...ICS 3000 Ion Chromatography System 346 Doc 065031 04 1 08...
Страница 368: ...ICS 3000 Ion Chromatography System 352 Doc 065031 04 1 08...
Страница 372: ...ICS 3000 Ion Chromatography System 356 Doc 065031 04 1 08...
Страница 374: ...ICS 3000 Ion Chromatography System 358 Doc 065031 04 1 08...