Connecting an Autosampler (Optional)
Doc. 065294-01 3/09
5
3. Route the sample outlet line
from the AS-DV through
the left side slot, at the front
of the IC system.
4. Connect the outlet sample
line from the AS-DV to port
S (5)
on the injection valve
Figure 5. IC System Injection Valve
3.2
Connecting an AS Autosampler
NOTE Dionex recommends using AS models that do not
have injection valves. If a valve is installed in the
autosampler, it is switched at the same time as the
valve in the IC system. This causes unnecessary
wear on the autosampler valve. If your AS has an
injection valve, contact Dionex for assistance.
1. Place the autosampler to the left of the IC system on the workbench.
2. Open the IC system and the autosampler doors.
3. Route the needle seal line that extends from the autosampler inject
port through the autosampler right side slot (see
through the IC system left side slot to the injection valve.
Figure 6. AS Autosampler Drip Tray and Right Side Slot
AS-DV
Sample
Outlet Line
Insert the waste line
from the IC system
injection valve here
Route the autosampler
needle seal line to the
injection valve inside
the IC system