ASE 150 Operator’s Manual
88
Doc. 065207-02 9/08
5.8.1
Removing the Static Valve from the System
1. Turn off the ASE 150 main power switch.
2. Follow the instructions in
to disconnect the gas source
and remove the right-side panel.
The static valve is installed on the lower left side of the component
panel, below the pressure relief valve (see
).
3. Using a 1/4-inch open-end wrench, disconnect the two Hastelloy fluid
lines from the valve
.
4. Disconnect the dark yellow
and orange
tubing from the press
fittings on the valve.
To disconnect a press fitting, use your fingers (or a small open-end
wrench) to press the ring on the fitting
in
, while at the same time
pulling the tubing
out
.
5. Using a 2.5-mm hex wrench, remove the four screws
that secure
the static valve to the ASE 150.
Figure 5-20. Removing the Static Valve
Mounting Screws (4)
Hastelloy
Lines (2)
1
4
2
1
4
2
3
Dark Yellow
3
Orange Tubing
1
1
1
2
1
3
1
4