ASE 150 Operator’s Manual
42
Doc. 065207-02 9/08
3.2
Running
3.2.1
Selecting the Method
1. Press
MENU
to display the
MENU
screen (see
2. Move the cursor to
SETUP
and press
ENTER
. The
SETUP
screen
appears (see
3. Press
ENTER
to move the cursor to the method editing field.
4. Press an arrow button to step through the custom method numbers
and the names of the preprogrammed methods. To edit a custom
method, see
. For descriptions of the preprogrammed
methods, refer to
.
5. When the method required for the run is displayed, press
ENTER
.
3.2.2
Selecting the Cell Size
The ASE 150 determines various internal operating parameters (for
example, the rinse volume) based on the cell size selected on the
SETUP
screen (see
). If the selected cell size does not match the size
of the cell in use, select the correct size:
1. On the
SETUP
screen, move the cursor to
CELL SIZE
and press
ENTER
.
2. Press an arrow button to step through the cell sizes.
3. When the correct cell size is displayed, press
ENTER
.
Figure 3-12. Menu Screen
Figure 3-13. Setup Screen
STATUS
SETUP
METHOD EDITOR
DIAGNOSTICS
METHOD
CELL SIZE
UNITS
OFF
PSI
100ML
PCB
REDUCE RELIEF
Содержание ASE 150
Страница 1: ...ASE 150 Accelerated Solvent Extractor Operator s Manual Document No 065207 Revision 02 September 2008...
Страница 32: ...ASE 150 Operator s Manual 24 Doc 065207 02 9 08...
Страница 64: ...ASE 150 Operator s Manual 56 Doc 065207 02 9 08...
Страница 76: ...ASE 150 Operator s Manual 68 Doc 065207 02 9 08...
Страница 114: ...ASE 150 Operator s Manual 106 Doc 065207 02 9 08...
Страница 132: ...ASE 150 Operator s Manual 124 Doc 065207 02 9 08...