NOTE The procedure for installing 30-mm filters in
stainless steel and zirconium cells is the same.
The photos in
show a zirconium cell.
Figure 3-5. Installing a 30-mm Cell Filter in a 34
-
, 66
-
, or 100
-
mL Cell
Insert a 30-mm
filter into the
cell at a slight
angle.
Position the filter
insertion tool
(P/N 056929)
over the filter.
Slowly push the
insertion tool
straight into the
cell.
Push down until
the filter is in full
contact with the
end cap.