2 • Description
Doc. 031842-01 5/02
2-13
2.4
Method Control
A method defines all of the timed events that occur during a sample extraction (see
Table 2-2). There are two types of methods:
_
Preprogrammed methods are application-specific methods created by Dionex.
These methods cannot be changed or deleted by the user. For a list of the
parameters for each preprogrammed method, see Section 2.5.
_
Custom methods are customer-programmable methods. All custom methods
initially contain the default method parameters. These methods can easily be
modified (i.e., customized) by the user as required. For instructions on how to
edit a custom method, see Section 3.2.
Parameter
Function
Value Range
TEMPERATURE
Temperature at which to heat the extraction cell.
Off, 40 to
200
°
C
(default=100)
STATIC TIME
Static solvent extraction time.
0 to 99 min
(default=5)
FLUSH
VOLUME
Amount of solvent to flush through the
extraction cell after the static heating step. The
FLUSH VOLUME
is expressed as a percentage
of the cell volume; for example, if the
FLUSH
VOLUME
is set to 50%, 5 mL is flushed through
a 10 mL cell, 17 mL is flushed through a 34 mL
cell, and so on.
0 to 150% vol
in 5% incre-
ments
(default=60)
PURGE TIME
Amount of time the cell is purged with nitrogen.
20 to 900 sec
(default=100)
STATIC CYCLE
Number of times the static heating and flushing
steps are performed. When more than one cycle
is specified, the flush volume is divided among
the cycles.
1 to 5
(default=1)
Table 2-2. Method Parameters
Содержание ASE 100
Страница 8: ...ASE 100 Accelerated Solvent Extractor vi Doc 031842 01 5 02...
Страница 24: ...ASE 100 Operator s Manual 2 10 Doc 031842 01 5 02 Figure 2 4 ASE 100 Rear Panel...
Страница 32: ...ASE 100 Operator s Manual 2 18 Doc 031842 01 5 02...
Страница 42: ...ASE 100 Operator s Manual 3 10 Doc 031842 01 5 02 Figure 3 2 Inspecting Cell O Rings...
Страница 60: ...ASE 100 Operator s Manual 3 28 Doc 031842 01 5 02...
Страница 69: ...4 Troubleshooting Doc 031842 01 5 02 4 9 Figure 4 1 ASE 100 Plumbing Diagram...
Страница 106: ...ASE 100 Operator s Manual 5 36 Doc 031842 01 5 02...
Страница 110: ...ASE 100 Operator s Manual A 4 Doc 031842 01 5 02...
Страница 113: ...B Installation Doc 031842 01 5 02 B 3 Figure B 1 ASE 100 Rear Panel...
Страница 150: ...ASE 100 Accelerated Solvent Extractor Index 8 Doc 031842 01 5 02...