2 • Description
Doc. 031842-01 5/02
2-11
2.3
Extraction Process
Before starting an extraction, perform the following steps. Refer to Chapter 3 for
detailed instructions for each step.
_
Prepare sample and load it into the extraction cell.
_
Set up a method.
_
Press the
START
button.
_
Verify that the system
STATUS
is
OVEN READY
(see Figure 2-5) and the LED
on the
START
button is flashing.
_
Place the collection bottle, with the bottle cap screwed on, in the holder.
Lower the needle mechanism.
_
Place the extraction cell in the cell holder and close the cell door.
_
Press
START
to begin the run.
From this point, the ASE 100 automatically performs the extraction process. An
extraction consists of six main steps:
_
Filling the cell with solvent
_
Heating the cell (equilibration)
_
Static extraction
_
Flushing with fresh solvent
_
Purging solvent from the system
_
End relief
Содержание ASE 100
Страница 8: ...ASE 100 Accelerated Solvent Extractor vi Doc 031842 01 5 02...
Страница 24: ...ASE 100 Operator s Manual 2 10 Doc 031842 01 5 02 Figure 2 4 ASE 100 Rear Panel...
Страница 32: ...ASE 100 Operator s Manual 2 18 Doc 031842 01 5 02...
Страница 42: ...ASE 100 Operator s Manual 3 10 Doc 031842 01 5 02 Figure 3 2 Inspecting Cell O Rings...
Страница 60: ...ASE 100 Operator s Manual 3 28 Doc 031842 01 5 02...
Страница 69: ...4 Troubleshooting Doc 031842 01 5 02 4 9 Figure 4 1 ASE 100 Plumbing Diagram...
Страница 106: ...ASE 100 Operator s Manual 5 36 Doc 031842 01 5 02...
Страница 110: ...ASE 100 Operator s Manual A 4 Doc 031842 01 5 02...
Страница 113: ...B Installation Doc 031842 01 5 02 B 3 Figure B 1 ASE 100 Rear Panel...
Страница 150: ...ASE 100 Accelerated Solvent Extractor Index 8 Doc 031842 01 5 02...