B • Installation
Doc. 031842-01 5/02
B-7
4. Compare the pressure readings on the
REGULATORS
screen with the
values listed in the table below. If a pressure reading does not meet
specification, contact Dionex.
Field
Function
Pressure Spec
SYSTEM
Pressure applied to the system
0.34
±
0.02 MPa
(50
±
3 psi)
COMPRESSION
Pressure applied when the oven
is compressed
0.90
±
0.03 MPa
(130
±
5 psi)
BOTTLE
Pressure applied to the solvent
bottle
0.03 to 0.07 MPa*
(5 to 10 psi)
AUTOSEAL
Pressure applied when the
AutoSeal air cylinder is
actuated
0.10
±
0.01 MPa
(15
±
2 psi)
*This pressure is nominally set for 0.04 MPa (6 psi). The pressure may
increase during periods of system inactivity, but will return to a value
within the normal range as soon as an extraction begins.
Содержание ASE 100
Страница 8: ...ASE 100 Accelerated Solvent Extractor vi Doc 031842 01 5 02...
Страница 24: ...ASE 100 Operator s Manual 2 10 Doc 031842 01 5 02 Figure 2 4 ASE 100 Rear Panel...
Страница 32: ...ASE 100 Operator s Manual 2 18 Doc 031842 01 5 02...
Страница 42: ...ASE 100 Operator s Manual 3 10 Doc 031842 01 5 02 Figure 3 2 Inspecting Cell O Rings...
Страница 60: ...ASE 100 Operator s Manual 3 28 Doc 031842 01 5 02...
Страница 69: ...4 Troubleshooting Doc 031842 01 5 02 4 9 Figure 4 1 ASE 100 Plumbing Diagram...
Страница 106: ...ASE 100 Operator s Manual 5 36 Doc 031842 01 5 02...
Страница 110: ...ASE 100 Operator s Manual A 4 Doc 031842 01 5 02...
Страница 113: ...B Installation Doc 031842 01 5 02 B 3 Figure B 1 ASE 100 Rear Panel...
Страница 150: ...ASE 100 Accelerated Solvent Extractor Index 8 Doc 031842 01 5 02...