Select the Method and Flush Syringe Speeds
Refer to the tables in
to select the correct method and flush
syringe speeds to ensure the backpressure generated does not exceed
690 kPa (100 psi).
•
If the AS is under Chromeleon or Chromeleon Xpress control, set the
syringe speeds on the Control panel or in a program. See
for details.
•
If the AS is
always
under front panel control (it is never connected to
Chromeleon or Chromeleon Xpress), set the syringe speeds on the
SYSTEM PARAMETERS
screen. See
Select Plumbing Configuration Options
1. On the
MODULE SETUP MENU
, press
3
to go to the
PLUMBING
CONFIGURATION
2. Move the cursor to the
SAMPLE SYRINGE VOLUME
field and select
the volume of the sample syringe installed.
NOTE For the simultaneous option, the sample syringe
size must be either 5,000 or 10,000
μ
L.
3. If the sample preparation option is installed, move the cursor to the
PREP SYRINGE VOLUME
field and select the volume of the prep
syringe.
4. Move the cursor to the
LOOP SIZE V1
field and enter the size of the
sample loop installed on the injection valve.
Figure B-44. Plumbing Configuration Screen Example
Help Prompt
PLUMBING CONFIGURATION
LOOP SIZE V1:
SAMPLE SYRINGE VOLUME:
PREP SYRINGE VOLUME:
SAMPLE SYRINGE CONFIG:
5
25
250
READY
uL
uL
mL
PREP SYRINGE CONFIG:
READY
uL
LOOP SIZE V2:
Содержание As
Страница 1: ...AS Autosampler Operator s Manual Document No 065051 Revision 03 January 2008...
Страница 190: ...AS Autosampler Operator s Manual 178 Doc 065051 03 1 08...
Страница 238: ...AS Autosampler Operator s Manual 226 Doc 065051 03 1 08...
Страница 338: ...AS Autosampler Operator s Manual 326 Doc 065051 03 1 08...
Страница 355: ...Index Doc 065051 03 1 08 Index 17 Z Z address 255...
Страница 356: ...AS Autosampler Operator s Manual Index 18 Doc 065051 03 1 08...