AS-HV Operator’s Manual
54
Doc. 065125-01 10/06
5. Using a cable tie, attach
the tubing through the
eyelet on the underside of
the housing (see
). Leave a loop
of about 10 mm (0.39 in)
in the cable tie.
Figure 5-5. Attaching the Cable Tie to the Tubing
6.
Locate one tube anchor along the X-axis cover, about 18 cm (7 in)
from the column case. Locate the second tube anchor toward the
back, about 11.4 cm (4.5 in) from the column case (see
7.
Attach the tubing to the middle tube anchor (the anchor installed
first), using the two tube markers (see
8.
Route the tubing along the side of the X-axis arm. Anchor the tubing
in position, spacing the tube anchors so as to minimize the route to the
IC (or other analyzer). When you finish, adjust the length of the
sample line to the analyzer.
Figure 5-6. Attaching the Tubing to the Middle Tube Anchor
Sampling Arm
Housing
Use a cable tie to
attach the tubing
through the eyelet
Install this tube
anchor first
X-Axis Cover
Install this tube
anchor second
Содержание AS-HV
Страница 1: ...AS HV Autosampler Operator s Manual Document No 065125 Revision 01 October 2006...
Страница 8: ...AS HV Operator s Manual vi Doc 065125 01 10 06...
Страница 28: ...AS HV Operator s Manual 20 Doc 065125 01 10 06...
Страница 54: ...AS HV Operator s Manual 46 Doc 065125 01 10 06...
Страница 66: ...AS HV Operator s Manual 58 Doc 065125 01 10 06...
Страница 92: ...AS HV Operator s Manual 84 Doc 065125 01 10 06...
Страница 116: ...AS HV Operator s Manual 108 Doc 065125 01 10 06...