![Dini Argeo DGT4X Скачать руководство пользователя страница 11](http://html1.mh-extra.com/html/dini-argeo/dgt4x/dgt4x_quick-start-manual_2505737011.webp)
11
QSG_ENG_DGT4X
Data
Register
DESCRIPTION
Gross weight
30001
Gross Weight value.
30002
Net weight
30003
Net Weight value.
30004
Input status
register
30005
Bit 15
(msb)
Bit 14
Bit 13
Bit 12
Bit 11
Bit 10
Bit 9
Bit 8
(lsb)
Active channel.
Active channel.
No function.
No function.
No function.
No function.
Status of input n. 2.
Status of input n. 1.
Bit 7
(msb)
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
(lsb)
1 = Scale unloaded (gross weight = 0).
Tare PT (1 = PT tare is active).
Tare (1 = Tare is active).
Overload condition (0 = No; 1 = Overload).
Underload condition (0 = No; 1 = Underload).
Weight Stability (0 = Unstable; 1 = Stable).
Gross Weight Polarity (0 = “+”; 1 = “-”).
Net Weight Polarity (0 = “+”; 1 = “-”).
Command status
register
30006
Last received command.
Bit 7
(msb)
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
(lsb)
Last command result.
Last command result.
Last command result.
Last command result.
Counting of processed commands.
Counting of processed commands.
Counting of processed commands.
Counting of processed commands.
Output status
register
30007
No Function.
Bit 7
(msb)
...
Bit 2
Bit 1
Bit 0
(lsb)
No function.
...
No function.
Digital output 2 status (0 = OFF; 1 = ON).
Digital output 1 status (0 = OFF; 1 = ON).
µV Channel 1
30111
µV value of the channel 1.
µV Channel 2
30112
µV value of the channel 2.
µV Channel 3
30113
µV value of the channel 3.
µV Channel 4
30114
µV value of the channel 4.
Bit 15 Bit 14
Active Channel
0
0
Channel 1
0
1
Channel 2
1
0
Channel 3
1
1
Channel 4
16. Modbus
16.1 MODBUS REGISTERS -
dep.CH
/
ind.ch
(1 SCALE)
This manual contains the main registers for reading data / sending commands.
Refer to the Modbus protocol manual for the complete list of available registers.