35
Optimized layout for A4 print.
tr 10
tr 8
tr 6
tr 4
tr 2
tr 1
tr1/2
tr1/4
TR no
1
2
3
4
5
6
7
8
9
0.trk.sp
5
0.traCk
4
DISAB
ENAB
888888
F.mode
A
funCt
1
rs.zero
5
USER_MAN_ENG_DGT1SP
LEGEND:
Indicates repeated pressing of
the key.
Parameter visible only
under certain conditions.
Parameter or menu subject
to approval.
Default value of the parameter.
Zero tracking time
10 divisions.
8 divisions.
6 divisions.
4 divisions.
2 divisions.
1 division.
1/2 division.
1/4 division.
Tracking disabled.
This menu allows to set zero tracking, i.e. the compensation parameter of the thermal drift of the scale; the set value corresponds to
the number of divisions that is reset to zero in the fixed time of 1 second.
Zero tracking
By pressing the key, or sending the zero command, the transmitter zeroes the gross weight on the scale. For a moment the display
shows “
zero
” and then it shows 0 (gross weight).
The semi-automatic zeroing cannot be performed if:
•
The weight on the scale is greater than the zero capacity (
0.perC
).
•
The weight is unstable.
Restoring zero
Semi-automatic zeroing
Zero restoring disabled.
Zero restoring enabled.
• If
auto.0
=
disab
:
The last zero in the memory before turning off the power is always
restored.
• If
auto.0
=
enab
/
CyCle
:
The last zero in the memory before turning off the power is restored
only if the auto-zeroing fails.
from 100 to 5000.
1000.
Once stability is reached, the transmitter waits for the time set in this parameter before
bringing the weight back to zero (if it is within the divisions set in
0.traCk
).
Содержание DGT1SP
Страница 1: ...www diniargeo com DGT1SP Digital weight transmitter ENGLISH USER MANUAL...
Страница 2: ......
Страница 6: ...6 Optimized layout for A4 print USER_MAN_ENG_DGT1SP...
Страница 57: ...57 Optimized layout for A4 print USER_MAN_ENG_DGT1SP Notes...
Страница 58: ...58 Optimized layout for A4 print USER_MAN_ENG_DGT1SP Notes...
Страница 59: ......