Pirani gauge tubes must be installed in a thread-down orientation in a clean, dry vacuum system.  While 
threading the gauge tube in to the manifold, the gauge tube cable should be disconnected to avoid 
damage.  In this way, twisting of the cable and the octal socket on the tube is avoided.  Care should be 
exercised to install the tubes in a dry part of the system. Since the instrument works on the principle of 
temperature rise, the probes will not work if they become filled with a liquid such as vacuum or diffusion 
pump oil.  The gauge tube should be protected against oil and other contaminants by installing it in such 
a way to protect it.  A good practice is to mount the gauge tube in the most vertically distant place from 
oil and other contaminants as applicable.  The gauge tube should be mounted in the most stable 
pressure region of the vessel to be measured.  For example, it would be better to install the gauge tube 
on a tank rather than on the pipe that is directly connected to a vacuum pump.  In the event of 
contamination, see section 6.0 for gauge tube cleaning instructions. 
 
 
 
The RS232 connection can be made to a PLC or computer via a male DB9 cable connection to the 
female jack on the DIGIVAC.  The DIGIVAC acts as a DCE, so a straight serial connection is 
appropriate. 
 
Please use the supplied 5V AC adapter with your Instrument.  This adapter provides clean short 
protected power to protect and insure accuracy of the internal circuitry. 
 
 

 
 
5.0   OPERATION. 

 
After installation, the DIGIVAC is ready for immediate operation.  The unit will normally provide accurate 
readings immediately; when used with a Pirani gauge however, occasionally a gauge tube will have 
absorbed material during storage and may require as much as 24 hours of operation before accurate 
readings are attained.  It is recommended that the DIGIVAC be energized continuously during vacuum 
system operation. In this way, the hot filament will not allow contaminants to condense. 
 
Only connect and disconnect cables with the power to the unit unplugged.  Make all connections to 
sensors, and relay outputs with the power disconnected.  NEVER DISCONNECT SENSOR OR 
OUTPUT WIRES WHILE UNIT IS POWERED UP. 
 
In cases where the system has contaminants, as is often the case with metalizing and coating 
equipment, it is often effective to isolate the gauge tube with a solenoid or manual valve during periods 
when contamination is most active. 

 

 

 

 
The Instrument has additional outputs which can be used: 
 

 

RS232 

– The instrument puts out a standard RS232 serial stream with settings 9600, 8, N, and 

1.  The unit transmits but does not receive, and displays the current vacuum indication in the 
current units. 

 

Analog out 

– This output reads from 0 to 5 Volts from a pressure of 1 micron all the way up to 5 

Torr.  There is a graduation of 1 milliVolt per millitorr.  Therefore, 10 milliVolts = 10 millitorr and 4 
Volts = 4 Torr. 

 

 

Содержание 200 Pirani

Страница 1: ...AC Model 200 Pirani Digital Vacuum Gauge Revsion K13A YOU MUST READ THIS MANUAL BEFORE USE Readings from 1 2000 millitorr The DIGIVAC Company 105B Church Street Matawan NJ 07747 www DIGIVAC com 732 76...

Страница 2: ...le of Operation 2 0 Construction 3 0 Unpacking and Inspection 4 0 Installation 5 0 Operation 6 0 Servicing and Calibration 7 0 Notes on Calibration 8 0 Understanding Microns Torr and Absolute Pressure...

Страница 3: ...rument the sensor configured the sensor cable the output interfaces and an international AC power adapter The instrument is housed in a rugged free standing vinyl clad metal enclosure It can be placed...

Страница 4: ...V AC adapter with your Instrument This adapter provides clean short protected power to protect and insure accuracy of the internal circuitry 5 0 OPERATION After installation the DIGIVAC is ready for i...

Страница 5: ...200 Surf over to http 192 168 0 200 or whatever your IP address was set to and say yes to security screens authorizing a DIGIVAC applet to run in your browser You should see a screen that looks like...

Страница 6: ...in Press carriage return or enter to submit the query or change the set point Note that while the web terminal is open the main 200 Pirani screen will not update If at any time the web terminal or th...

Страница 7: ...at the instrument is in fact incorrect In many cases the problem will be with a tube that is fouled or a system that is operating improperly It is recommended that a spare tube be kept on hand and sto...

Страница 8: ...altitude It also varies widely with local weather conditions Variations of one half inch in a single day are common The word vacuum means pressure lower than atmospheric or suction but in describing...

Страница 9: ...closure For instruments that will be used in the field particularly in cryogenic applications a padded shoulder strap case is available COMPATIBILITY WITH OTHER GAUGE TUBES On special order DIGIVAC I...

Страница 10: ...CULAR PURPOSE NON INFRINGEMENT OR ANY OTHER WARRANTY EXPRESSED OR IMPLIED LIMITATION OF LIABILITY You agree and acknowledge DIGIVAC shall have no liability to you whatsoever for any direct indirect pu...

Отзывы: