background image

product name:

product model:

working procedure:

Calibration Operation Instruction

1、adjust ACA0
  Set tap position in AC10A,adjust potentiometer VR1,let LCD display 
0.000-0.001A                                                               
2,Check datum mark:                                                     
Set tap position in ACV,input 100mV(60Hz) by use standard ACV current 
source ,let LCD display 0.999V~1.001V    
Test indication:

PICTURE INSTRUCTION

OPERATION STEPS

Digital Multimeter

QM-1549

ACV

ACA0 potentiometer VR4

ACV adjust potentiometer 
VR2

Measuring range   Calibrating spot   Normal display range(50-400HZ)       
400mV                  0mV             0.0-4.9mV(hand movement range)    
400mV                 100mV            99.2-100.8mV(hand movement range) 
400mV                 100mV            0.095-0.105V(automatic range)     
4V                     1V              0.988~1.012V                      
 
40V                  10.00V            9.88~10.12V                      
400V                 100.0V            99.4~101.6V                       
  600V                 600V              587~613V 
square wave.triangular wave,amplitude limiting wave test                  
Measuring range   Calibrating spot   Normal display range(50-400HZ)       
40V                  35.00V            35.40~34.60V