Digicare Biomedical Technology Inc.
Operators Manual Pulse Oximeter Monitor DIGIOXI “PO-930”
4
SECTION 2 - Technical specification
A
.
Mechanical Description
Size
: 94x254x254mm
Weight
: 3,5 Kg
Color
: White and Black
B
.
Power Requirements
AC voltage Input
: 120VAC; 60Hz (220VAC and/or 50Hz Optional)
Internal Battery
: 12 VDC/2,9Ah Sealed
Battery Duration
: 7 Hours.
Recharge Time
: 18 Hours in use on AC line.
C
.
SpO
2
- Pulse Oximeter
SpO
2
Range
: 0% to 100% adult/pediatric/neonate
SpO
2
Accuracy
:
±
2% (70 - 100%),
±
3% (50 - 69%) unspecified (0 - 49%)
Pulse Rate Range
: 30 to 250 BPM
Pulse Rate Accuracy
: ± 2%
Pulse Tone
: Pitch adjusted by SpO
2
Sensor Types
: Finger, Universal, Wrap, Disposable.
D
.
Heart Rate
Heart Rate Range
: 30 to 255 BPM
Heart Rate Accuracy
: 2%
E
.
Display
Type
: High Intensity CFL-LCD
Dot Matrix
: 240x64 Dots
Effective
Display
Area
: 178 mm x 69 mm x 25mm
Viewing
Angle
: 150°