Operator’s Manual
Страница 1: ...DIGIOXI DIGIOXI Model PO 930 PULSE OXIMETER MONITOR Operator s Manual...
Страница 2: ...Digicare Biomedical Technology Inc Operators Manual Pulse Oximeter Monitor DIGIOXI PO 930 2 Quality and Technology To Touch Life...
Страница 3: ...ment Specifications 5 H RS232 output Optional 5 SECTION 3 Controls Connectors and Indicators A Front Panel 6 B Rear Panel 7 C Monitoring Mode Screen 8 D Alarms Indicators 8 SECTION 4 Operation Instruc...
Страница 4: ...ried out by persons authorized by the manufacturer It is up to the user to ensure that any applicable regulations respecting the installation and operation of the monitor be observed The operator shou...
Страница 5: ...e Oximeter Pulse Rate PLETHYSMOGRAM Waveform The DIGIOXI PO 930 is intended for use by persons trained in professional health care The operator must be thoroughly familiar with the instructions in thi...
Страница 6: ...h Sealed Battery Duration 7 Hours Recharge Time 18 Hours in use on AC line C SpO2 Pulse Oximeter SpO2 Range 0 to 100 adult pediatric neonate SpO2 Accuracy 2 70 100 3 50 69 unspecified 0 49 Pulse Rate...
Страница 7: ...ent Specifications Temperature Operating 19 c to 35 c Storage 4 5 c to 43 5 c Relative Humidity Operating 20 80 Non Condensing Storage 10 90 Non Condensing H RS232 OUTPUT Optional Serial Output to con...
Страница 8: ...U L S E DIGIOXI PO 930 Pulse Oximeter L C D C O N T R A S T 01 02 03 04 05 04 06 07 08 09 10 01 AC indicator 02 ON OFF Switch 03 Trends ON OFF Push button 04 Increment and decrement alarm limits and...
Страница 9: ...CAUTION Equipment must be grounded at all times to avoid dangerous electrical shock hazard WARNING For continued protectionagainst fire hazard replace only with same type andrating of fuse DANGER Exp...
Страница 10: ...THYSMOGRAM WAVEFORM 03 SpO2 Indicates the SpO2 value LOW and HIGH alarm limits 04 PULSE Indicates the Pulse Rate value LOW and HIGH limits D Alarm Indicators ON AC OFF ALARM SELE CT TRENDS ON OFF S p...
Страница 11: ...the monitorization screen 5 At power on the monitor assume standards values for the following items Audible Alarm Two minutes OFF Pulse Rate High Limit 140 BPM Pulse Rate Low Limit 60 BPM SpO2 High L...
Страница 12: ...d not be displayed Alarm Sound Enabled Press the or push button to increase or decree the alarm sound F Trends ON OFF ON AC OFF TRENDS ON OFF S p O 2 P U L S E DIGIOXI PO 930 Pulse Oximeter L C D C O...
Страница 13: ...reduced hemoglobin and the absorption due to all other tissue constituents The problem is solved with a two wavelength pulsatile system The pulsation of arterial blood flow present at a particular te...
Страница 14: ...ries These are available from DIGICARE using the following part number 1 Finger Probe 10 cord P N PO 710 2 Finger Probe 1 5 cord P N PO 711 3 Ear Probe P N PO 730 4 Disposable Wrap Probe P N PO 724 5...
Страница 15: ...nemia hypothermia arterial occlusion to the sensor or cardiac arrest Inaccurate measurements may be caused by Incorrect application of dysfunctional hemoglobin s such as carboxyhemoglobin or methemogl...
Страница 16: ...g instability in SpO2 monitoring Check the probe installation site perfusion and body temperature SEARCHING This message is displayed when the monitor is self adjusting the ideal point of signal monit...
Страница 17: ...ruments brushes rough surfaced materials or make contact with anything that can scratch the panel B Probes The probes are the only surfaces of this monitor that come in contact with the patient Clean...
Страница 18: ...e during the warranty period if returned to the factory with prior authorization transportation prepaid Not covered by this agreement are repairs necessitated by any of the following conditions 1 Inad...
Страница 19: ...gy Inc Operators Manual Pulse Oximeter Monitor DIGIOXI PO 930 17 Date Equip Purch ________________________ Expiration Date ___________________ DIGICARE BIOMEDICAL TECHNOLOGY INC 107 Commerce Road Boyn...